Evaluation of precision and trueness of intraoral scanners for newborns with cleft lip and palate: 3D analysis of digital models.
Juliana-Theresa Schell, Bert Braumann, Teresa Kruse, Sarah Achterrath, Tamara N Shamlian, Isabelle Graf
Department of Orthodontics and Dentofacial Orthopedics, Medical Faculty and University Hospital of Cologne, Kerpener Str. 32, Cologne, 50931, Germany. · Private Practice, Fresno, CA, USA.
Key points
From the paper's conclusion- Under in-vitro conditions, all IOS digitized neonatal maxillary anatomy with high technical accuracy. Trueness and precision depended on the combination of device and cleft morphology rather than on morphological complexity alone, but all IOS captured the relevant morphological features. Clinical feasibility in neonates was not assessed and remains to be established in in-vivo studies.
Abstract
BACKGROUND: Children born with cleft lip and palate (CLP) require intensive interdisciplinary care from birth onward. Intraoral scanners (IOS) have emerged as a promising alternative for neonatal maxillary data acquisition. This study evaluated the precision and trueness of five commonly used IOS using a standardized landmark-based methodology complemented by full-surface deviation analysis, applied to neonatal maxillary models with increasing morphological complexity. METHODS: Three neonatal plaster casts were digitized: An edentulous maxilla without cleft, a unilateral and a bilateral cleft maxilla. Ten pyramidal landmarks defining five transversal distances were constructed along the alveolar ridge. These resulting digital master models (DMM) were 3D-printed (Formlabs Form 2, 25 μm) and their dimensional fidelity independently verified by digital caliper (± 0.01 mm). Each printed model was scanned ten times by the same operator using Primescan, CS3600, Trios 3, iTero Element 2 and Medit i500. Linear trueness and precision were quantified in Rhinoceros 3D. Full-surface trueness was assessed in Geomagic Control 3D using a standardized alignment protocol. Statistical analysis was performed using the Kruskal-Wallis test and Mann-Whitney U post-hoc tests (p < 0.05). RESULTS: Trueness varied between IOS and depended on the combination of device and cleft morphology. Mean signed linear deviations were minimal in the non-cleft model (all scanners within ± 0.03 mm) and reached - 0.147 mm for Medit in the bilateral cleft model, where Primescan and Trios remained essentially unbiased in the non-cleft and bilateral cleft models. Between-scanner differences were significant in both cleft models (p < 0.001). Intra-scanner reliability, an aspect of precision, ranged from ICC 0.832 to 1.000. Full-surface trueness differed significantly between scanners in all three models (p < 0.01). Mean RMS deviations remained below 0.07 mm for four of the five devices in every model, whereas iTero reached 0.224 mm in the bilateral cleft model. Compared with the non-cleft model, in-tolerance percentages decreased in both cleft models, most markedly in posterior and undercut regions. CONCLUSION: Under in-vitro conditions, all IOS digitized neonatal maxillary anatomy with high technical accuracy. Trueness and precision depended on the combination of device and cleft morphology rather than on morphological complexity alone, but all IOS captured the relevant morphological features. Clinical feasibility in neonates was not assessed and remains to be established in in-vivo studies. CLINICAL TRIAL NUMBER: Not applicable.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Trueness and precision of 5 intraoral scanners under different conditions in arches: A 3D comparative in vitro analysis.Dental and medical problems · 2026
- Effect of Preparation Design on the Trueness and Precision of Intraoral Scanners: A Comparative Analysis Across Two IOS Systems.Journal of esthetic and restorative dentistry : official publication of the American Academy of Esthetic Dentistry ... [et al.] · 2026
- Scanner-Dependent Influence of Palatal Surface Acquisition on Trueness, Precision and Transverse Arch Reproduction During Maxillary Complete-Arch Intraoral Scanning.Journal of dentistry · 2026
- Effect of scan calibrator morphologies on the accuracy of calibrated complete-arch implant intraoral scans: an in vitro study.Journal of dentistry · 2026
- STL Triangle Count and Trueness in Prepared Tooth Regions: Effects of Scanner System and Scanning Mode Across Intraoral, Laboratory, and Industrial Scanners.Journal of dentistry · 2026
- Limitations of RMS error in evaluating intraoral scanner accuracy: Comparison with linear and angular distortions from an enhanced metrology protocol.The Journal of prosthetic dentistry · 2026