Evaluation of bone-conducted sound during ear surgery: an in vivo pilot study.
Katarina Gille, Mario Fleischer, Marcus Neudert, Thomas Zahnert, Matthias Bornitz
Department of Otorhinolaryngology Head and Neck Surgery, Faculty of Medicine and University Hospital Carl Gustav Carus, TU-Dresden, Dresden, Germany. katarina.gille@ukdd.de. · Department of Otorhinolaryngology Head and Neck Surgery, Faculty of Medicine and University Hospital Carl Gustav Carus, TU-Dresden, Dresden, Germany.
Key points
From the paper's conclusion- Based on these findings, standardized noise monitoring is recommended as a quality assurance measure during otological surgery.
Abstract
The impact of noise induced by drilling on hearing during ear operations has been discussed several times in the past. However, direct measurements of the equivalent sound pressure level in patients undergoing ear surgery have not previously been reported. This pilot study presents an easy-to-use method based on a calibrated accelerometer to determine the equivalent sound pressure level potentially applicable in the clinical routine. During seven monitored cochlear implant surgeries, the equivalent A-weighted sound pressure level LAF ranged from 111 to 126 dB(A)eqSPL at its maximum, whereas C-weighted peak levels LCpeak showed peaks up to 140 dB(C)eqSPL. The recommended daily noise exposure limit was exceeded for three of the seven patients, suggesting the potential for drilling-induced temporal shifts in hearing threshold. Based on these findings, standardized noise monitoring is recommended as a quality assurance measure during otological surgery.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Noninvasive indirect intracochlear pressure monitoring during cochlear implantationHearing Research · 2026
- Quantifying Soft‐Surgery in Cochlear Implantation: Multimodal Data From 30 International SpecialistsThe Laryngoscope · 2026
- A Prospective, Multicenter Clinical Investigation on the Safety and Performance of a New Active Transcutaneous Bone-Anchored Implant SystemOtology & Neurotology · 2025
- EP9.23 Intraoperative Noise Exposure Across the Spectrum of Hip Surgery: Hip Arthroscopy, Periacetabular Osteotomy, and Total Hip ArthroplastyJournal of Hip Preservation Surgery · 2026
- Operating room noise levels in anaesthesia a systematic review and meta-analysis.British journal of anaesthesia · 2026
- High-Precision Insertion Force Measurements for Cochlear Implant Surgery: Accuracy Analysis of an Improved DeviceCurrent Directions in Biomedical Engineering · 2026