Evaluation of computer-aided detection products in community and facility tuberculosis screening: 2025 policy update.
Mikashmi Kohli, Aurelien Mace, Cecily Miller, Sam Linsen, Dennis Falzon, Matthew Arentz, Rigveda Kadam, Sandra Vivian Kik,
FIND, Geneva, Switzerland. · World Health Organization, Geneva, Switzerland.
Key points
From the paper's conclusion- These findings demonstrate that CAD performance varies markedly across products, underscoring the need for standardised, product-specific evaluation and careful threshold selection. This work constituted evidence informing the WHO policy update on CAD use for TB screening.
Abstract
BACKGROUND: Since the World Health Organization (WHO) issued its initial recommendation on computer-aided detection (CAD) use in 2021, the number of commercially available products has expanded substantially, necessitating an updated independent evaluation to inform global policy. METHODS: We conducted a standardised, head-to-head evaluation of eight commercially available CAD products for adult tuberculosis (TB) screening using two large, well-characterised, geographically diverse chest radiography (CXR) libraries representing community-based and facility-based screening use cases. CAD products were selected through a WHO public call for expressions of interest and evaluated independently on a secure validation platform with offline installation. Diagnostic accuracy was assessed using receiver operating characteristic (ROC) analysis, partial area under the curve (AUC) in clinically relevant specificity ranges (40-60%), and comparisons with expert radiologist performance and WHO target product profile thresholds. RESULTS: Diagnostic performance varied substantially across products. In community screening, AUCs ranged from 0.72 to 0.84, and in facility screening from 0.77 to 0.86. Six CAD products (qXR, Insight CXR, CAD4TB, InferRead DR Chest, Genki Edge and DrAid) demonstrated optimal diagnostic accuracy as they had overlapping confidence intervals with the expert radiologists' estimates and met WHO performance considerations for screening use, whereas two products (Radify Chest and JVIEWER-X) underperformed. At a fixed sensitivity of 90%, specificities ranged from 26-56% in community screening and 30-49% in facility screening. CONCLUSIONS: These findings demonstrate that CAD performance varies markedly across products, underscoring the need for standardised, product-specific evaluation and careful threshold selection. This work constituted evidence informing the WHO policy update on CAD use for TB screening.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy and operational performance of computer-aided detection software use cases for TB screening: a retrospective comparative evaluation in Ho Chi Minh City, Viet Nam.BMJ public health · 2026
- Beyond overall accuracy: subgroup performance and threshold calibration of AI-assisted chest radiography for equity-oriented pulmonary tuberculosis screening.Frontiers in medicine · 2026
- Artificial intelligence for pulmonary tuberculosis detection on chest X-ray images: A systematic review and diagnostic accuracy meta-analysismedRxiv · 2026
- Diagnostic accuracy of AI-assisted chest radiographs in tuberculosis screening: A Ghanaian clinical studyPLoS ONE · 2026
- Comparative cost analysis of community-based tuberculosis screening: Computer-Aided Detection (CAD) alone versus CAD with point-of-care C-reactive protein testing in South Africa and Lesotho.PLOS global public health · 2026
- Computer-aided detection pragmatic threshold calibration in TB active case finding: A mixed-methods study in Tondo, Manila.PLOS global public health · 2026