Accuracy and Treatment Experience of a Next-Generation Continuous Glucose Monitoring System in Children With Diabetes: A Clinical Trial.
Iman S Al-Gadi, Ahlam Alotaibi, Hessah Alotaibi, Sharifah Al Issa, Amal Aljuhani, Reem A Al Khalifah
Division of Pediatric Endocrinology, Department of Pediatrics, College of Medicine, King Saud University, Riyadh, Saudi Arabia. · Division of Pediatric Endocrinology, Department of Pediatrics, Maternity and Children Hospital, Al-Madinah Al-Munawwarah, Saudi Arabia.
Key points
From the paper's conclusion- The Syai CGM system showed good accuracy with SMBG and a favorable user experience in this pediatric cohort.
Abstract
AIM OF STUDY: To assess the diagnostic accuracy between the next-generation, factory-calibrated continuous glucose monitoring (CGM) system and self-monitored blood glucose (SMBG) measurements, and to evaluate user experience with the system in Saudi children with diabetes. METHODS: We performed a diagnostic accuracy study of the next-generation Synchronized Artificial Intelligence System CGM system (Syai) more than 60 days. Sixty children aged 0 to 18 years with diabetes requiring insulin treatment wore the sensor on the upper arm, abdomen, thigh, or buttock for up to 14 days, followed by rotation to another site. Sensor performance was tested against SMBG. We reported the mean absolute relative difference (MARD) between CGM and SMBG measurement pairs, the proportion of CGM values meeting the 15/15, 20/20, and 30/30 agreement criteria. RESULTS: The mean age of the children was 9.0 ± 4.4 years, and 30 (50.0%) were men. The average diabetes duration was 2.7 ± 2.8 years, mean hemoglobin A1c was 9.1 ± 1.8%, and time in range (70-180 mg/dL) was 50.5 ± 19.6%. A total of 238 sensors were worn for 60 days, providing 2364 paired CGM-SMBG measurements for agreement analysis from 52 participants, including 1480 pairs from the upper arm, 585 from the abdomen, and 299 from the buttock/thigh. The overall pair-weighted MARD was 11.74% (participant-cluster bootstrap 95% CI, 10.79%-13.05%). The overall 15/15, 20/20, and 30/30 agreement rates were 76.44%, 86.29%, and 94.92%, respectively. No serious adverse events were reported. CONCLUSION: The Syai CGM system showed good accuracy with SMBG and a favorable user experience in this pediatric cohort. CLINICAL TRIAL REGISTRATION: Saudi Food and Drug Authority (SFDA) Clinical Trials Registry, http://www.sfda.gov.sa/en/node/3549755.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy, safety, and usability of a real-time continuous glucose monitoring system in pediatric diabetes management.Journal of diabetes investigation · 2026
- Performance of a 15-Day Continuous Glucose Monitoring System in Children and Adolescents With Diabetes.Journal of diabetes science and technology · 2026
- Clinical accuracy of Dexcom G6 and G7 continuous glucose monitors in hospitalized pediatric patients with type 1 diabetes: a real-world studyAnnals of Pediatric Endocrinology & Metabolism · 2026
- Performance of the i-SENS CareSens Air CGM System in Compliance with the European CGM (eCGM) Clinical Performance Criteria: A Pooled Analysis.Diabetes therapy : research, treatment and education of diabetes and related disorders · 2026
- Pediatric Inpatient Continuous Glucose Monitor Accuracy by Sensor Wear Day, Hospital Day, and Glucose Rate of Change.Diabetes technology & therapeutics · 2026
- Glycaemic outcomes of three months of continuous glucose monitoring in children, adolescents and young adults with type 1 diabetes at a Kenyan tertiary hospital: a single-arm prospective study.medRxiv · 2026