Performance of a 15-Day Continuous Glucose Monitoring System in Children and Adolescents With Diabetes.
Manuela Link, Christian Denzer, Torben Biester, Jantje Weiskorn, Julian Ziegler, Franziska Liebrich, Désirée Dunstheimer, Jakob Krämer,
Institut für Diabetes-Technologie Ulm GmbH, Germany. · Division of Pediatric Endocrinology and Diabetology, Department of Pediatrics and Adolescent Medicine, University Medical Center Ulm, Germany.
Key points
From the paper's conclusion- The observed performance and safety of CSAir in children and adolescents are similar to those in adults. These data may support an extension of regulatory approval of CSAir for children aged 2 years and older.
Abstract
BACKGROUND: The CareSens Air (CSAir) continuous glucose monitoring system is approved for use in adults in South Korea and received adults-only CE marking in early 2024. The aim of this prospective, multicenter, pivotal study in children and adolescents with insulin-dependent diabetes mellitus was to determine if the performance and safety of CSAir in these populations are comparable to those observed in adults. METHOD: Sensors were worn for 15 days on the upper arm in children and adolescents aged 7 to 17 years and on the upper arm and/or upper buttock in children aged 2 to 6 years. Frequent capillary comparator measurements were performed during in-clinic sessions at the beginning, middle, and end of the sensor-wearing period. Point accuracy was assessed via agreement rates (AR) and mean absolute relative difference (MARD). Additional assessments included clinical point accuracy, technical reliability, and safety. RESULTS: For subjects aged 7 to 17 years, %20/20 AR was 90.8%, and MARD was 8.9%. A total of 99.9% of values fell within Zones A and B of the Diabetes Technology Society error grid. Sensor survival probability reached 75.2%. In children aged 2 to 6 years, %20/20 ARs were 86.6% and 98.4%, and MARDs were 10.3% and 7.8% for arm- and buttock-worn sensors, respectively. No serious adverse events were reported. Mild and moderate adverse device effects comprised skin issues such as itching or erythema at the sensor application or insertion site. CONCLUSIONS: The observed performance and safety of CSAir in children and adolescents are similar to those in adults. These data may support an extension of regulatory approval of CSAir for children aged 2 years and older.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy and Treatment Experience of a Next-Generation Continuous Glucose Monitoring System in Children With Diabetes: A Clinical Trial.Journal of diabetes science and technology · 2026
- Performance of the i-SENS CareSens Air CGM System in Compliance with the European CGM (eCGM) Clinical Performance Criteria: A Pooled Analysis.Diabetes therapy : research, treatment and education of diabetes and related disorders · 2026
- Accuracy, safety, and usability of a real-time continuous glucose monitoring system in pediatric diabetes management.Journal of diabetes investigation · 2026
- Accuracy and Safety of the LinX Continuous Glucose Monitoring System in Adults With Diabetes: A Prospective Multicentre Open-Label Clinical Performance Study.Diabetes, obesity & metabolism · 2026
- Clinical accuracy of Dexcom G6 and G7 continuous glucose monitors in hospitalized pediatric patients with type 1 diabetes: a real-world studyAnnals of Pediatric Endocrinology & Metabolism · 2026
- Pediatric Inpatient Continuous Glucose Monitor Accuracy by Sensor Wear Day, Hospital Day, and Glucose Rate of Change.Diabetes technology & therapeutics · 2026