Extensive metallosis following reverse shoulder arthroplasty managed with custom glenoid reconstruction: A case report.
Tarishi Parmar, Mohammad Daher, Miguel Fiandeiro, Peter Boufadel, Mohamad Fares, Joseph A Abboud
Division of Shoulder and Elbow Surgery, Rothman Orthopaedic Institute, Philadelphia, PA, USA.
Key points
From the paper's conclusion- Catastrophic metallosis secondary to modular junction failure may require staged revision and patient-specific reconstruction when extensive bone loss precludes standard revision implants.
Abstract
BACKGROUND: Metallosis is a rare but serious complication following reverse shoulder arthroplasty (RSA) that can lead to progressive bone loss and implant failure. Although revision strategies have been described planned two-stage revision with custom implants has not previously been reported. CASE PRESENTATION: A 55-year-old female presented with pain and dysfunction following RSA. Examination demonstrated severe motion limitation, while imaging revealed loosening and progressive bone loss. Intraoperatively, we identified extensive metallosis with micromotion at the modular humeral junction, and a large glenoid defect. The initial glenoid baseplate was also positioned relatively high, resulting in impingement and subsequent scapular notching, which may have contributed to the metallosis. Owing to extensive metallosis, severe osteolysis, poor bone stock, and uncertain infection status, a staged revision was performed. An antibiotic spacer was placed and glenoid was bone grafted. Eventually definitive reconstruction was performed using a custom glenoid baseplate and humeral component because standard revision implants could not achieve stable fixation. At 1 year, the patient had improved shoulder function with forward elevation of 130° and was able to perform activities of daily living. CONCLUSION: Catastrophic metallosis secondary to modular junction failure may require staged revision and patient-specific reconstruction when extensive bone loss precludes standard revision implants.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Combined Glenoid Osteotomy and Bone Block Augmentation for Severe Posterior Shoulder Dysplasia and Instability: A Case Report.JBJS case connector · 2026
- Salvage reverse shoulder arthroplasty after failed plate fixation of proximal humeral fractures: Clinical outcomes, complications, and prognostic factors.Shoulder & elbow · 2026
- The «SPA-CE» classification for glenoid bone loss in reverse shoulder arthroplasty: a CT-based analysis of 211 shoulders.Musculoskeletal surgery · 2026
- Polyethylene glenospheres in reverse total shoulder arthroplasty have greater long-term implant survivorship than metallic glenospheres: A New Zealand Joint Registry Study.Journal of shoulder and elbow surgery · 2026
- Reverse shoulder arthroplasty after failed tendon transfer: A systematic review of clinical outcomes and complications from the FP-UCBM shoulder study group.Shoulder & elbow · 2026
- [Translated article] Clinical and radiological outcomes when using standard versus retentive polyethylene humeral cups in Delta III reverse shoulder arthroplasty.Revista espanola de cirugia ortopedica y traumatologia · 2026