Warm, ventilate, stabilize: Clinical feasibility and non-inferiority study of MINAR™ infant warmer resuscitator.
Reza Widianto Sudjud, Jenifer Kiem Aviani, Lia Nazliah Abdullah
Department of Anaesthesiology and Intensive Care, Faculty of Medicine, Hasan Sadikin General Hospital, Universitas Padjadjaran, Bandung, West Java, Indonesia. · Department of Paediatrics, Bandung Kiwari General Regional Hospital, Bandung, West Java, Indonesia.
Key points
From the paper's conclusion- MINAR™ demonstrated feasibility, safety, and non-inferior clinical performance compared with the Dräger Resuscitaire® for neonatal postnatal stabilization. The observed advantages in oxygenation and respiratory distress remained independent of baseline respiratory severity, supporting further evaluation in larger multicenter randomized trials.
Abstract
BACKGROUND: Effective neonatal resuscitation requires integrated thermal support, respiratory assistance, and continuous physiological monitoring. MINAR™, a locally manufactured infant warmer resuscitator developed in Indonesia, was designed to support neonatal stabilization. This study evaluated the clinical feasibility, safety, and non-inferiority of MINAR™ compared with the Dräger Resuscitaire®. METHODS: This single-center, randomized, non-inferiority feasibility trial allocated live-born neonates requiring postnatal resuscitation or stabilization in a 1:1 ratio to MINAR™ or Dräger Resuscitaire®. Primary outcomes included respiratory rate (RR), heart rate (HR), peripheral oxygen saturation (SpO2), body temperature, Apgar score, and Downes score measured at baseline and 5, 10, 15, and 60 min. Minimum clinically important differences were predefined. Repeated-measures ANCOVA was performed to assess the potential confounding effect of baseline Downes score. RESULTS: Forty-eight neonates were enrolled, and 38 were included in the final analysis (MINAR™, n = 18; Dräger Resuscitaire®, n = 20). Baseline characteristics were comparable except for a higher baseline Downes score in the Dräger group (p = 0.004). MINAR™ demonstrated non-inferior performance across all predefined outcomes with no differences exceeding the non-inferiority margins. After adjustment for baseline respiratory distress, MINAR™ remained associated with faster improvement in oxygen saturation and greater reduction in Downes score, whereas RR, HR, body temperature, and Apgar score showed comparable longitudinal trajectories between groups. No device-related adverse events or malfunctions were observed. CONCLUSIONS: MINAR™ demonstrated feasibility, safety, and non-inferior clinical performance compared with the Dräger Resuscitaire® for neonatal postnatal stabilization. The observed advantages in oxygenation and respiratory distress remained independent of baseline respiratory severity, supporting further evaluation in larger multicenter randomized trials. CLINICAL TRIAL REGISTRY: ISRCTN88734666.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Initial CPAP of 5 versus 8 cmH2O during delivery room stabilisation of very preterm infants: the PHILODENDROOM pilot randomised trial.European journal of pediatrics · 2026
- Variations in neonatal resuscitation practice and interventions: an international survey study.Neonatology · 2026
- Time-critical oxygen support: Clinical outcomes according to initial oxygen support device in early respiratory failurePneumologia · 2026
- Donor Human Milk Implementation in a High-Mother's-Own-Milk NICU: Impact on Feeding Progression and Clinical Outcomes in Extremely Low Birth Weight Infants.Nutrients · 2026
- Impact of real-time respiratory function monitoring on neonatal mask ventilation training: a multicentre simulation-based crossover study.BMJ paediatrics open · 2026
- Targeted Oxygen for Initial Resuscitation of Preterm Infants: The TORPIDO 30/60 Randomized Clinical Trial.JAMA · 2026