Investigation of S-Detect AI (a computer-aided diagnosis tool) versus ultrasound operators for thyroid nodule classification using British Thyroid Association guidelines: A pilot diagnostic accuracy study.
Kieran Walker, James Michael, Mark Ballard, Geoffrey Chilvers, Steve Colley, Khalid Hussain, Lloyd Rickard
University Hospitals Birmingham NHS Foundation Trust, UK. · Birmingham City University, UK.
Key points
From the paper's conclusion- Operators achieved superior diagnostic performance. S-Detect's high false-negative rate appears largely driven by over-reliance on Doppler. Its clinical utility is limited within this single-centre study. Future iterations excluding Doppler would likely achieve substantially improved sensitivity.
Abstract
INTRODUCTION: Thyroid nodules affect more than 60% of the population, approximately 5% being malignant. The British Thyroid Association U-classification guides ultrasound risk stratification across the United Kingdom but is subject to inter-observer variability. Artificial intelligence decision support tools may offer more standardised assessment, yet no peer-reviewed study has validated S-Detect, a computer-aided diagnosis tool integrated into Samsung ultrasound systems, against the British Thyroid Association framework. This study compares S-Detect with experienced ultrasound operators using histology as the reference standard. METHODS: This retrospective pilot diagnostic accuracy study assessed 32 thyroid nodules (September 2023 to August 2024). S-Detect served as the index test applied retrospectively on a Samsung RS85 Prestige platform; prospectively assigned U-scores were the comparator and surgical histology the reference standard. A secondary analysis varied Doppler input across all four options in the 11 malignancies misclassified by S-Detect to test whether Doppler selection contributed to misclassification. RESULTS: Operators achieved superior balanced performance (sensitivity: 78.9%, specificity: 84.6%, accuracy: 81.3%, and area under the curve: 0.870) compared with S-Detect (42.1%, 100%, 65.6%, area under the curve: 0.729), which under-classified 58% of malignancies. Inter-method agreement was slight (κ = 0.175), with significant directional discordance (McNemar p = 0.008). Doppler manipulation produced substantial variability: 81.8% of misclassified cases showed U-score change attributable solely to Doppler selection, with peripheral patterns yielding zero correct malignancy classifications. CONCLUSION: Operators achieved superior diagnostic performance. S-Detect's high false-negative rate appears largely driven by over-reliance on Doppler. Its clinical utility is limited within this single-centre study. Future iterations excluding Doppler would likely achieve substantially improved sensitivity.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Experience-dependent integration of a commercial AI system with C-TIRADS for thyroid nodule assessment: a retrospective single-center study.Frontiers in endocrinology · 2026
- Diagnostic performance and clinical integration of an artificial intelligence-assisted ultrasound decision support tool for thyroid nodules at a tertiary cancer centre.Ultrasound (Leeds, England) · 2026
- Real-Time AI-Based ACR TI-RADS Classification of Thyroid Nodules on the Ultrasound Console: A Prospective Agreement Study in a Region with Historical Iodine Deficiency.Diagnostics (Basel, Switzerland) · 2026
- Evaluating Diagnostic Strategies for Thyroid Nodule Assessment Using S-Detect in Transverse and Longitudinal Ultrasound Planes.Technology in cancer research & treatment · 2026
- Artificial Intelligence in Thyroid Nodules and Cancer: Clinical Validation and Real-World Performance.Endocrine connections · 2026
- Artificial intelligence-based systems as auxiliary tools for thyroid nodule malignancy detection on ultrasound: a systematic review and meta-analysis of diagnostic accuracy.Gland surgery · 2026