Determinants of Continuous Glucose Monitoring Accuracy in Critically Ill Patients Requiring Mechanical Ventilation and Vasopressor Support.
Ola Friman, Samuel Wiqvist, Marcus Lind, Pia Zetterqvist, Olav Rooijackers, Anders Oldner, Johan Mårtensson
Department of Physiology and Pharmacology, Section of Anaesthesia and Intensive Care, Karolinska Institutet, Stockholm, Sweden. · Department of Learning, Informatics, Management and Ethics, Division of Medical Statistics, Karolinska Institutet, Stockholm, Sweden.
Key points
From the paper's conclusion- In critically ill patients receiving mechanical ventilation and vasopressor therapy, CGM accuracy improves substantially with increasing sensor wear time and is independently associated with arterial pH. Early postinsertion sensor instability and acid-base disturbances appear to be important determinants of CGM performance in critical illness.
Abstract
BACKGROUND: Continuous glucose monitoring (CGM) accuracy in critically ill patients may be influenced by altered physiology and concurrent therapies. We assessed independent associations between sensor wear time, arterial blood glucose (ABG), noradrenaline-equivalent (NE) dose, erythrocyte volume fraction, pH, lactate, body-mass index and dialysis to CGM performance (Dexcom G6®) in this setting. METHODS: Secondary analysis of a prospective observational study including 40 critically ill adults receiving intravenous insulin, mechanical ventilation and vasopressor therapy. Paired ABG and CGM values (n = 2946) were analyzed. The primary outcome was percentage absolute relative difference (ARD). Secondary outcomes were the proportions of CGM readings meeting ISO 15197:2013 and CLSI POCT12-A3 criteria. Multivariable generalized linear mixed models with patient-level random intercepts were fitted, applying B-splines for continuous covariates. RESULTS: Median (IQR) ARD was 10.8% (5.2-18.8) (mean ARD 12.7 [95% CI 10.7-15.3] %); 64.5% and 56.0% of readings met ISO and CLSI criteria, respectively. Sensor wear time (P < 0.001) and pH (P ≤ 0.02) were independently associated with ARD and both categorical accuracy metrics. Predicted ARD decreased, and the probability of meeting ISO/CLSI criteria increased, during the first 50 h after sensor insertion and then stabilized. The pH-accuracy relationships were approximately linear; in linear-term sensitivity analyses, each 0.1-unit increase in pH was associated with a 1.5 percentage-point higher ARD (P = 0.009) and 32% lower odds of meeting CLSI criteria (P = 0.001). NE-dose (P = 0.006) and ABG (P < 0.001) were independently associated with ISO-defined accuracy. Spline analyses suggested reduced accuracy at higher NE-doses (>0.6 µg/kg/min) and lower ABG (<5 mmol/L [<90 mg/dL]), although observations were limited in these ranges. CONCLUSIONS: In critically ill patients receiving mechanical ventilation and vasopressor therapy, CGM accuracy improves substantially with increasing sensor wear time and is independently associated with arterial pH. Early postinsertion sensor instability and acid-base disturbances appear to be important determinants of CGM performance in critical illness.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Continuous glucose monitoring for diabetes mellitus in the perioperative and intensive care setting: a prospective observational validation study✰.British journal of anaesthesia · 2026
- Accuracy of an Interstitial Continuous Glucose Management System in Critically Ill Patients: A Prospective Observational Study.Journal of diabetes science and technology · 2026
- Clinical validation of continuous glucose monitoring in glycemic management of critically ill sepsis patients: A prospective randomized clinical study.The Journal of international medical research · 2026
- Pediatric Inpatient Continuous Glucose Monitor Accuracy by Sensor Wear Day, Hospital Day, and Glucose Rate of Change.Diabetes technology & therapeutics · 2026
- Association of continuous glucose monitoring with in-hospital adverse outcomes among critically ill patients: a real-world propensity score-matched study.Frontiers in endocrinology · 2026
- Clinical accuracy of Dexcom G6 and G7 continuous glucose monitors in hospitalized pediatric patients with type 1 diabetes: a real-world studyAnnals of Pediatric Endocrinology & Metabolism · 2026