Diabetes technology & therapeutics

Continuous Glucose Monitoring and Automated Insulin Delivery in Infants with Type 1 Diabetes: Real-World Evidence to Support Future International Guidelines.

Marco Marigliano, Andrea E Scaramuzza, Claudia Arnaldi, Riccardo Bonfanti, Valentino Cherubini, Roberto Franceschi, Fortunato Lombardo, Giulio Maltoni,

Department of Surgery, Dentistry, Pediatrics, Section of Pediatric Diabetes and Metabolism, Verona, Italy. · Pediatric Diabetes, Endocrinology and Nutrition, Pediatric Unit, Ospedale Maggiore, ASST Cremona, Cremona, Italy.

Guideline

Key points

From the paper's conclusion
  • This real-world evidence supports incorporating AID consideration into future international guidelines from the time of pediatric diagnosis.

Abstract

To evaluate efficacy and safety of continuous glucose monitoring (CGM) and automated insulin delivery (AID) in infants. A retrospective multicenter study (13 centers) of 76 children diagnosed with antibody-positive type 1 diabetes before age 2. Outcomes at 12 months were compared between AID users (n = 47) and multiple daily injections (MDI)/SAP users (n = 29). AID systems included Medtronic 780 G (38%), Tandem Control-IQ (38%), and CamAPS FX (24%). CGM median wear time was 97.0% (93.3; 98.1). AID users achieved significantly higher TIR (65% vs. 56%, P = 0.041) and time in tight range (TITR) (42% vs 32.5%, P = 0.038). In multivariable regression, AID was the sole independent predictor of lower HbA1c (P = 0.035). Zero severe hypoglycemia (SH) events occurred in the AID group (vs. 7.1% in MDI/SAP). Use of CGM is safe in very young children, with AID systems proving superior to traditional therapy. This real-world evidence supports incorporating AID consideration into future international guidelines from the time of pediatric diagnosis.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers