Aseptic loosening of small hydroxyapatite-coated quadrangular tapered femoral stems: camouflage in the Canadian Joint Replacement Registry.
Christiaan H Righolt, Colton Poitras, Rohit Bansal, Colin D Burnell, Thomas R Turgeon, Eric R Bohm
Orthopaedic Innovation Centre, Winnipeg, MB, Canada. · Department of Surgery, University of Manitoba, Winnipeg, MB, Canada.
Key points
From the paper's conclusion- The increased risk of revision of these small stems camouflages a highly increased risk of revision for aseptic loosening in undersized stems (small stem/big cup). This study was unable to ascertain whether this applies to other quadrangular tapered stems. Due to the small number of affected patients, it is important to repeat this study in other jurisdictions. Surgeons should be skeptical if a smaller size stem seems appropriate during broaching than preoperative templating or cup size suggests, as this may increase the risk of aseptic loosening.
Abstract
BACKGROUND: Small cementless hydroxyapatite-coated quadrangular tapered femoral stems are at increased risk of revision, but it is unclear whether this holds for appropriately-sized and undersized stems alike. This study aimed to assess the association between stem size and revisions for aseptic loosening. METHODS: Primary total hip arthroplasties with an Avenir (1547; Zimmer-Biomet), Corail (16,893; DePuy Synthes), or Polarstem (3403; Smith+Nephew) stem during 2012-2023 were identified from the Canadian Joint Replacement Registry. Cox proportional hazards models were used to estimate the association between stem/cup size and revision for aseptic loosening while adjusting for age, sex, and implant characteristics. Standardised stem sizing from A-K (smallest-largest) was utilised. RESULTS: Between different stems, the proportion of size A was 1.2-5.0% and of size B 3.6-12.2%. ⅓ (21/63) of aseptic loosening revisions occurred for size A/B. Increased risk of revision for aseptic loosening for size A stems was found with a hazard ratio (HR) of 46 (95% CI, 9-226) when paired with ⩾56-mm cups, 47 (14-157) when paired with 52-54-mm cups, and 17 (4-78) when paired with ⩽50-mm cups (reference size C+ with ⩾56-mm cups). Size B stems were also at increased risk of aseptic loosening, with an HR of 15 (95% CI, 4-52) and 9 (95% CI, 3-32) for ⩾56 and 52-54-mm cups respectively. CONCLUSIONS: The increased risk of revision of these small stems camouflages a highly increased risk of revision for aseptic loosening in undersized stems (small stem/big cup). This study was unable to ascertain whether this applies to other quadrangular tapered stems. Due to the small number of affected patients, it is important to repeat this study in other jurisdictions. Surgeons should be skeptical if a smaller size stem seems appropriate during broaching than preoperative templating or cup size suggests, as this may increase the risk of aseptic loosening.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Association between cementless femoral stem geometry and early aseptic loosening following primary total hip arthroplasty.Hip international : the journal of clinical and experimental research on hip pathology and therapy · 2026
- Collared Stems Are Associated With Reduced Early Aseptic Loosening and All-Cause Revision Rate.The Journal of arthroplasty · 2026
- Periprosthetic strain-adaptive bone remodeling of conservative metaphyseal fitting versus conventional anatomic cementless femoral stem.Hip international : the journal of clinical and experimental research on hip pathology and therapy · 2026
- Increased Risk of Periprosthetic Femoral Fracture with Cementless Collared Metadiaphyseal-Filling Stems Compared to Cemented Fixation for Femoral Neck Fractures in Patients Aged 65 Years or Older: An American Joint Replacement Registry Analysis.The Journal of arthroplasty · 2026
- Collared, Uncemented Stems Markedly Reduce Periprosthetic Femoral Fracture Risk in Women ≥ 70 Years: A Swiss National Registry Study.The Journal of arthroplasty · 2026
- Early Periprosthetic Femoral Fractures and Stem Subsidence After Direct Anterior Total Hip Arthroplasty Using a Collared, Fully Hydroxyapatite-Coated Stem.Journal of orthopaedics · 2026