Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine

Bio-inspired electrosurgical knife with lubricant-assisted porous-groove surface and micro-serrated edge for reducing tissue adhesion and thermal injury.

Chen Li, Lifangke Liu, Jing Zhao, Zhen Shi, Lin Wang, Liping Zhou, Xiaodong Wang, Weibo Dou

School of Mechanical and Electrical Engineering, Shaanxi University of Science and Technology, Xi'an, Shaanxi, China. · Shaanxi Province Geriatric Hospital, Xi'an, China.

Key points

From the paper's conclusion
  • These results demonstrate an effective engineering strategy for improving the anti-adhesion and thermal-safety performance of electrosurgical knives.

Abstract

Tissue adhesion and thermal injury remain major challenges in high-frequency electrosurgery because carbonized residues can accumulate on the electrode surface and reduce cutting stability. To address this issue, a coating-free bio-inspired electrosurgical knife was designed and fabricated from 316L stainless steel by selective laser melting. Inspired by loach-skin lubrication and anti-fouling characteristics, a porous-groove blade surface was designed for lubricant storage and transport. Inspired by crab-chela serrations, a micro-serrated cutting edge was introduced to reduce cutting resistance. Fluid-flow simulation showed that the porous-groove surface promoted lubricant retention and local recirculation, reducing near-wall resistance by up to 20.73%. Friction tests indicated that Dp = 350 μm and Lp = 800 μm produced the lowest lubricated friction coefficient of 0.362 among the tested porous-groove surfaces. Static cutting simulation and cutting-force tests confirmed that the micro-serrated edge reduced cutting resistance, with the lowest peak cutting force of approximately 0.605 N obtained at H1 = 300 μm. Powered electrosurgical tests showed that the sample with Dp = 350 μm, Lp = 800 μm, and H1 = 300 μm achieved the lowest apparent adhesion mass of 2.07 ± 0.67 mg, corresponding to a 61.5% reduction compared with the DT401 commercial smooth control electrode. The thermal damage areas on the upper and lateral tissue surfaces were reduced by 56.75% and 54.44%, respectively. These results demonstrate an effective engineering strategy for improving the anti-adhesion and thermal-safety performance of electrosurgical knives.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Bio-inspired electrosurgical knife with lubricant-assisted porous-groove surface and micro-serrated edge for reducing tissue adhesion and thermal injury — Science Skim