In Vitro Evaluation of Pulsed Thulium:YAG and High-Power Ho:YAG Lasers with Multiple Pulse Modulations.
Alejandra Bravo-Balado, Antoni Sánchez-Puy, Paula Izquierdo, Pietro Diana, Alberto Piana, Damiano Stracci, Paola Arena, Sílvia Gràcia Garcia,
Department of Surgery and Morphological Sciences, Universitat Autònoma de Barcelona, Barcelona, Spain. · Department of Urology, Stone Unit, Fundació Puigvert, Barcelona, Spain.
Key points
From the paper's conclusion- p-Tm:YAG and Ho:YAG lasers demonstrated distinct ablation morphologies depending on pulse modulation and laser settings. At intermediate and low-energy settings, p-Tm:YAG (LP) produced wider, shallower fissures, whereas several Ho:YAG modalities achieved greater AV through deeper fissures. Ho:YAG (VB™) showed the most similar ablation morphology to p-Tm:YAG (LP). Whether these differences translate into clinical performance remains to be determined.
Abstract
OBJECTIVE: To compare stone ablation volume (AV) and ablation morphology between pulsed thulium:YAG (p-Tm:YAG) and high-power holmium:YAG (Ho:YAG) lasers with different pulse modulation technologies using an in vitro stone model. MATERIALS AND METHODS: BegoStone phantoms (15:6 powder-to-water ratio) were used. A motorized arm applied the laser on the phantoms at a constant speed (0.7 mm/s). Three laser settings were tested: 1.5 J × 5 Hz, 1 J × 20 Hz, and 0.3 J × 50 Hz. For p-Tm:YAG, short pulse (SP) and long pulse (LP) modes were evaluated; for Ho:YAG, SP, LP, Vapor Tunnel™ (VT™), and Virtual Basket™ (VB™) modes were tested. Five 21-mm fissures were created per setting. AV was calculated from fissure width (WOF) and depth (DOF) measured by optical microscopy. RESULTS: A total of 555 measurements were analyzed. At 1.5 J × 5 Hz, p-Tm:YAG (SP) produced the deepest fissures (DOF = 1000 µm vs 523.3-734.3 µm, p < 0.05) and was the only modality with a width-to-depth ratio <1, indicating deeper and narrower fissures. In contrast, p-Tm:YAG (LP) showed DOF and width-to-depth ratios comparable with Ho:YAG. At 1 J × 20 Hz, p-Tm:YAG (LP) generated wider, shallower fissures with a lower AV (11.9 mm³ vs 16.9-19.7 mm³, p < 0.05), whereas Ho:YAG (VB™) demonstrated a comparable AV. At 0.3 J × 50 Hz, p-Tm:YAG (LP) and Ho:YAG (VB™) exhibited similar width-to-depth ratios (1.3-1.4), whereas Ho:YAG (LP) and (VT™) achieved greater AV (7.4-8.0 vs 3.4 mm³; p < 0.05), driven by greater DOF. CONCLUSIONS: p-Tm:YAG and Ho:YAG lasers demonstrated distinct ablation morphologies depending on pulse modulation and laser settings. At intermediate and low-energy settings, p-Tm:YAG (LP) produced wider, shallower fissures, whereas several Ho:YAG modalities achieved greater AV through deeper fissures. Ho:YAG (VB™) showed the most similar ablation morphology to p-Tm:YAG (LP). Whether these differences translate into clinical performance remains to be determined.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- When long pulse is not enough: inferior dusting efficiency of pulsed thulium:YAG compared with Ho:YAG and thulium fiber laser in a standardized in vitro renal model.World journal of urology · 2026
- In vitro ablation rates of Ho:YAG, p-Tm:YAG and TFL lasersWorld Journal of Urology · 2025
- An in-vitro comparative study of magneto holmium: YAG and fiber dust thulium fiber lasers-primary effects on ablation, retropulsion, fragmentation, and temperature changes using phantom urologic stones.World journal of urology · 2026