Platform differences: Retropulsion, fragmentation, and dust in pulsed thulium:YAG lithotripsy.
Takashi Yoshida, Kazuyoshi Nakao, Yusuke Shirasu, Yusuke Motoki, Takahiro Nakamoto, Gen Kawa
Department of Urology, Saiseikai Noe Hospital, Osaka, Japan. · Department of Urology and Andrology, Kansai Medical University, Osaka, Japan.
Key points
From the paper's conclusion- Despite sharing the same pulsed Thulium:yttrium-aluminium-garnet laser medium, the two platforms exhibited distinct laser-stone interaction profiles. Further clinical evaluation is warranted to determine the relevance of these differences.
Abstract
OBJECTIVE: Our objective was to compare bubble appearance, retropulsion kinetics, and lithotripsy efficiency between two pulsed Thulium:yttrium-aluminium-garnet laser platforms with distinct engineering characteristics, including pumping configuration and peak-power capability. MATERIALS AND METHODS: We compared the RevoLix HTL (LISA Laser Products GmbH) and the Thulio (Dornier MedTech GmbH) across six settings (0.2-1.0 J, 10-50 Hz, short/long pulse). Standardized 5 × 5 × 5 mm artificial stones on a V-shaped rail and calcium oxalate dihydrate dust from one clinical case were evaluated. Imaging was used to qualitatively assess bubbles, and Kinovea and a custom Python pipeline quantified the retropulsion and dust particle area. An equal-weight composite score summarized stone weight loss and retropulsion. RESULTS: Qualitative imaging demonstrated that the RevoLix HTL produced well-defined bubble outlines, whereas the Thulio generated less defined outlines. The Thulio showed pulsatile retropulsion, whereas the RevoLix HTL exhibited smoother displacement and lower mean retropulsion at all six settings (e.g., 1.0 J/10 Hz long pulse: mean ± standard deviation [SD] 3.1 ± 0.6 vs. 15.3 ± 2.1 mm; mean difference -12.1 mm [95% confidence interval -15.3 to -8.9]). In fixed-stone tests, the Thulio achieved greater stone weight loss at 0.4 J/25 Hz long pulse (mean ± SD 17.5 ± 5.0 vs. 35.0 ± 5.8 mg; mean difference -17.5 mg [95% confidence interval -26.9 to -8.1]). At equal weighting, RevoLix HTL had a higher composite score at five of six settings, although the ranking was weighting dependent. Exploratory particle analysis showed smaller particle areas with RevoLix HTL at all three settings. CONCLUSION: Despite sharing the same pulsed Thulium:yttrium-aluminium-garnet laser medium, the two platforms exhibited distinct laser-stone interaction profiles. Further clinical evaluation is warranted to determine the relevance of these differences.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- An in-vitro comparative study of magneto holmium: YAG and fiber dust thulium fiber lasers-primary effects on ablation, retropulsion, fragmentation, and temperature changes using phantom urologic stones.World journal of urology · 2026
- When long pulse is not enough: inferior dusting efficiency of pulsed thulium:YAG compared with Ho:YAG and thulium fiber laser in a standardized in vitro renal model.World journal of urology · 2026
- In vitro ablation rates of Ho:YAG, p-Tm:YAG and TFL lasersWorld Journal of Urology · 2025
- Smart Lithotripsy: Lasers, Energy Sources and Efficiency.Current urology reports · 2026
- Comparative Laser Efficacy and Clinical Outcomes of Pulsed Thulium: YAG versus standard Holmium: YAG Lithotripsy.Urologia internationalis · 2026
- In Vitro Evaluation of Pulsed Thulium:YAG and High-Power Ho:YAG Lasers with Multiple Pulse Modulations.Journal of endourology · 2026