Haemodynamic effects of vena cava occlusion at different occlusion ratios and sites: A mock circulatory loop study.
Guangmao Liu, Jianqiang Hao, Zhaohui Zhang, Tianshuo Liu, Shengshou Hu
State Key Laboratory of Cardiovascular Disease, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Key points
From the paper's conclusion- SVC occlusion and IVC occlusion below the renal veins reduce cardiac preload and increase renal venous (RVn) flow, without significantly reducing CO. However, compared with SVC occlusion, IVC occlusion below the renal veins requires a higher occlusion ratio and provides limited ventricular unloading. IVC occlusion below the right atrium is not recommended, as this would lead to a significant reduction in CO and have adverse effects on the circulatory system.
Abstract
BACKGROUND: The haemodynamic effects of vena cava occlusion are related to the patient's preserved cardiac function. Investigating the haemodynamic effects is critical for the clinical translation of vena cava occlusion devices. METHODS: A mock circulatory loop (MCL) system was developed for the in vitro haemodynamic simulation of the systemic venous system. Haemodynamic effects of vena cava occlusion were assessed under different heart failure states, different occlusion ratios and distinct occlusion sites. RESULTS: Superior vena cava (SVC) occlusion significantly reduced both central venous pressure (CVP) and renal venous pressure (RVnP) while causing only a limited decrease in cardiac output (CO), however, jugular venous pressure (JVP) increased significantly. Inferior vena cava (IVC) occlusion below the right atrium caused negative CVP, a significant reduction in CO and increased RVnP. In contrast, IVC occlusion below the renal veins preserved CO while reducing RVnP and improving renal perfusion. Moreover, effective pressure gradients were observed at occlusion ratios of 82.3%-87.1% for SVC occlusion and IVC occlusion below the right atrium, compared with 94.5%-96.1% for IVC occlusion below the renal veins. CONCLUSION: SVC occlusion and IVC occlusion below the renal veins reduce cardiac preload and increase renal venous (RVn) flow, without significantly reducing CO. However, compared with SVC occlusion, IVC occlusion below the renal veins requires a higher occlusion ratio and provides limited ventricular unloading. IVC occlusion below the right atrium is not recommended, as this would lead to a significant reduction in CO and have adverse effects on the circulatory system.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Mock Circulatory Loop With Autoregulated Renal Circulation Based on Perfusion Pressure for Circulatory Support Devices Evaluation.Artificial organs · 2026
- Cannula Pairing Determines Distal Limb Perfusion During Peripheral Veno-Arterial Extracorporeal Membrane Oxygenation: An Experimental Mock-Loop Study.Artificial organs · 2026
- Venous dimension of shock: Integrating arterial inflow and venous back-pressure in hemodynamic assessment.World journal of critical care medicine · 2026
- Hemodynamic Effects of Three Different Mechanical Circulatory Support Strategies in Isolated Acute Right Heart Failure.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- Pressure behavior in a roller pump-driven low-flow VACC mock circuit with inactive de-airing pump: A descriptive bench characterization under reduced patient-side effective filling pressure conditions.PloS one · 2026
- Haemodynamic effects of selective venoconstriction with centhaquine in spontaneously breathing patients with mixed hypovolaemic-vasodilatory shock.Experimental physiology · 2026