Intravascular hemolysis and clinical outcomes in continuous-flow LVAD patients: A single-center retrospective analysis.
Ali Onder Kilic, Cagatay Engin, Tahir Yagdi, Mustafa Ozbaran
Department of Cardiovascular Surgery, University of Health Sciences, Bursa Medical School, Bursa City Hospital, Bursa, Turkey. · Department of Cardiovascular Surgery, Ege University Medical School, Izmir, Turkey.
Key points
From the paper's conclusion- Baseline LDH alone is not a reliable predictor of pump thrombosis. In contrast, a ⩾4-fold dynamic rise in LDH relative to the patient's own baseline provides excellent negative predictive value. This dynamic surge serves as a practical, device-independent decision-support tool to rule out pump thrombosis with high confidence in all cf-LVAD patients.
Abstract
OBJECTIVE: To determine the incidence of intravascular hemolysis in patients implanted with continuous-flow left ventricular assist devices (LVADs), to describe longitudinal changes in hemolysis-related laboratory markers (LDH, total/direct bilirubin, INR), and to assess their diagnostic value for pump thrombosis and cerebrovascular events. METHODS: We retrospectively analyzed 100 end-stage heart failure patients (2014-2015). To ensure consistent longitudinal analysis, 59 long-term survivors (42 HeartWare HVAD, 17 HeartMate II) were included. For each patient, a dynamic LDH increase ratio (peak/preoperative baseline) was calculated. The diagnostic performance of the criterion "LDH >250 U/L and ⩾4-fold increase from baseline" was evaluated. Cox regression, ROC analysis, and Kaplan-Meier survival analysis were performed. RESULTS: During 1-year follow-up, 17 patients experienced thrombotic or cerebrovascular events (8 confirmed pump thrombosis). While absolute LDH values at fixed time points were not discriminative, the dynamic LDH increase ratio was significantly higher in the thrombosis group (p = 0.0102), with an AUC of 0.840. Preoperative LDH was a significant risk factor in Cox regression (HR: 1.005, p = 0.009), reflecting baseline disease severity. The ⩾4-fold LDH increase criterion demonstrated a Negative Predictive Value (NPV) of 97.6%, indicating excellent rule-out performance. No significant difference was found between device types (p > 0.05), supporting a device-independent biomarker dynamic. CONCLUSION: Baseline LDH alone is not a reliable predictor of pump thrombosis. In contrast, a ⩾4-fold dynamic rise in LDH relative to the patient's own baseline provides excellent negative predictive value. This dynamic surge serves as a practical, device-independent decision-support tool to rule out pump thrombosis with high confidence in all cf-LVAD patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Predicting Survival Beyond 5 Years On Continuous-Flow LVAD: A Landmark Analysis From The STELLAR Registry.ESC heart failure · 2026
- Impact of a durable left ventricular assist device on inflammation and cardiac remodelling in subjects with advanced heart failure.Open heart · 2026
- Comparing hemolysis among extracorporeal pumps: A large-scale retrospective study.Perfusion · 2026
- Intraoperative Goal-Directed Perfusion Improves Outcomes in Left Ventricular Assist Device Explant Heart Transplantation.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- Pre-implant right heart catheterization and liver biomarkers in left ventricular assist device candidatesEuropean Heart Journal · 2026
- Hemoadsorption combined with hemodialysis versus hemodialysis alone on mortality in end-stage kidney disease: a randomized, open-label, multicenter trialNature Communications · 2026