A feasibility pilot study of a resource-adapted ECMO-based ex vivo heart perfusion platform.
Serhii Sudakevych, Sofia Chaikovska, Stepan Maruniak, Gavrylo Kovtun, Bohdan Hulitskyi, Igor Kuzmych, Lawrence A Renna, Teryn Roberts,
Department of Anesthesiology and Extracorporeal Treatment, Heart Institute of the Ministry of Health of Ukraine, Kyiv, Ukraine. · Autonomous Reanimation and Evacuation (AREVA) Research Institute and Innovation Center, The Geneva Foundation, San Antonio, TX, USA.
Key points
From the paper's conclusion- These findings demonstrate the technical feasibility of the platform and support its potential as an accessible alternative for prolonged normothermic heart perfusion in resource-limited settings.
Abstract
BackgroundCold ischemic injury remains a major limitation in heart transplantation, particularly when donor organs must be transported over long distances. Normothermic ex vivo heart perfusion may reduce ischemic damage and enable functional assessment of the organ before implantation; however, the cost, system complexity, and logistical requirements of commercial systems limit broad implementation outside well-resourced transplant programs.MethodsA compact, resource-adapted ex vivo heart perfusion device (Ukrainian patent UA 156631 U) was evaluated during normothermic perfusion of three explanted human hearts at 34°C. Perfusion targets included aortic inflow of 700-800 mL/min, with epinephrine titrated as needed. Arterial perfusate was sampled at baseline and hourly to assess acid-base status, electrolytes, glucose, lactate, hemoglobin, and myocardial injury markers. Heart weight was recorded before and after perfusion, and myocardial tissue underwent semi-quantitative histopathologic evaluation. Given the exploratory design, results were summarized descriptively as individual case values over time.ResultsAcross all three experiments, normothermic ex vivo perfusion was associated with progressive metabolic stabilization. pH increased from baseline acidosis (6.95-7.00) to near-physiologic values by 180 minutes (7.30-7.33). Uncorrected lactate values declined to <5 mmol/L by 180 minutes in all cases, while glucose decreased steadily and electrolytes remained within narrow ranges. Myocardial injury markers increased after reperfusion, with the steepest rise during the first hour, followed by slower increases thereafter. Heart weight increased modestly (3.53%-4.58%), consistent with mild edema, and histopathology showed predominantly mild-to-moderate changes. No air embolism or device-related mechanical complications were observed.ConclusionIn this early three-case feasibility study, a resource-adapted ECMO-based platform successfully maintained aortic inflow and supported ex situ perfusion of explanted human hearts for 4-6 hours, while enabling continuous metabolic assessment. These findings demonstrate the technical feasibility of the platform and support its potential as an accessible alternative for prolonged normothermic heart perfusion in resource-limited settings. Further controlled studies incorporating direct aortic root pressure monitoring, standardized functional assessment, and transplantation endpoints are warranted to validate its performance and clinical applicability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Ex Vivo Heart Perfusion Versus Cold Storage in Prolonged Preservation Period: A Juvenile Porcine Experimental Model.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- "Heart in a Box," a Novel Organ Care System for Heart Transplantation.Cardiology in review · 2026
- From Ice to Perfusion: A Narrative Review of Modern Heart Preservation Systems.Journal of clinical medicine · 2026
- Extending Normothermic Ex Situ Heart Preservation and Viability to 48 Hours via Protein Supplementation and Plasma Exchange.ASAIO journal (American Society for Artificial Internal Organs : 1992) · 2026
- Ex Situ Echocardiography During Normothermic Perfusion: Serial Physiologic Full Cardiac-Output Assessments for Transplant Suitability.The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation · 2026
- Redox Phenotyping of the Donor Heart: A New Paradigm for Precision Heart Transplantation.Antioxidants (Basel, Switzerland) · 2026