The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation

Ex Situ Echocardiography During Normothermic Perfusion: Serial Physiologic Full Cardiac-Output Assessments for Transplant Suitability.

Wyeth D Alexander, Takahiro Nakashima, Brandon A Hernandez, Karen G Zimmerman, Sarah D King, S Sikandar Raza, Carol Ling, Alvaro Rojas-Peña,

University of Michigan Medical School Extracorporeal Life Support Laboratory, 1150 Medical Center Dr, B560 MSRB2, Ann Arbor, Michigan, 48109, USA. · University of Michigan Medical School Department of Anesthesiology, Frankel Cardiovascular Center, 1425 E. Ann Street, Floor 4, Ann Arbor, Michigan, 48109, USA.

Key points

From the paper's conclusion
  • Physiologic FCO guideline-directed ESE should facilitate determination of transplant suitability in currently discarded donor hearts and thereby increase the number available for transplantation.

Abstract

BACKGROUND: Clinical normothermic ex situ donor-heart perfusion (NEHP) uses retrograde aortic infusion which prevents hemodynamic assessment. We sought to optimize and validate guideline-directed ex situ echocardiography (ESE) for assessment of transplant suitability using a novel full-cardiac output (FCO) circuit. METHODS: Following transthoracic echocardiography (TTE), porcine hearts were procured and maintained on NEHP for 24 hours. At hours 1, 12, and 24, maintenance NEHP was transitioned to low-preload working-heart mode (LPWH). Although not physiologic, LPWH ensured optimal coronary perfusion and allowed ESE. Hearts were transitioned to physiologic FCO mode and ESE was repeated. NEHP was returned to maintenance. LPWH-ESE and FCO-ESE parameters were compared to determine their ability to stratify function. RESULTS: Forty-two echocardiograms (6 TTE, 36 ESE) were performed (n=6; 55±8 kg; female 50%; hearts 311±40gm). When compared to FCO-ESE, LPWH-ESE at hour 1 underestimated function (cardiac output (CO) 0.78±0.38 vs 3.26±1.70L/min, p=0.012; cardiac index (CI) 0.0024±0.0010 vs 0.0102±0.0048L/m/gm, p=0.009; stroke volume (SV) 8.0±3.5 vs 30.0±8.2mL, p=0.001; stroke work (SW) 279±102 vs 1683±1134mL/mmHg, p=0.026) and correlation was poor (R2 0.22-0.36). FCO-ESE successfully stratified function between hours 1 and 12 (LVEF 54.6±13.0 vs 33.6±6.0%, p=0.011; CO 3.26±1.70 vs 1.37±0.86L/min, p=0.003; CI 0.0102±0.0048 vs 0.0053±0.0027L/m/gm, p=0.016; SV 30.0±8.2 vs 15.3±6.6mL, p=0.001; SW 1683±1134 vs 686±472mL/mmHg, p=0.029), hours 12 and 24 (LVEF 33.6±6.0 vs 53.0±8.2%, p=0.036, and hours 1 and 24 (SV 30.0±8.2 vs 23.1±6.6mL, p=0.032). LPWH-ESE could not. CONCLUSION: Physiologic FCO guideline-directed ESE should facilitate determination of transplant suitability in currently discarded donor hearts and thereby increase the number available for transplantation. SOCIAL MEDIA: Guideline-directed Ex Situ Echocardiography during physiologic full cardiac output normothermic ex situ perfusion can reproducibly assess donor heart suitability for transplantation #EchoFirst @ASE360 @DHDrake.