Average glandular dose of patients undergoing digital breast tomosynthesis mammography examination at East Avenue Medical Center, Philippines.
E S Mones, R C F Caballar, J P Cruz
The Graduate School, University of Santo Tomas, Manila, Philippines 1008; e-mail: everjoymones@gmail.com. · Department of Radiological Sciences, East Avenue Medical Center, Quezon City, Philippines 1100; e-mail: julie_p_cruz@yahoo.com.
Key points
From the paper's conclusion- The study aims to analyse the average glandular dose (AGD) received by symptomatic patients during their mammography examination conducted in TomoHD mode at East Avenue Medical Center (EAMC).
Abstract
Digital breast tomosynthesis (DBT) is a medical imaging modality utilised in acquiring multiple two-dimensional images in various projections to visualise a three-dimensional image of the breast. The study aims to analyse the average glandular dose (AGD) received by symptomatic patients during their mammography examination conducted in TomoHD mode at East Avenue Medical Center (EAMC). Routine performance tests based on the American College of Radiology (ACR) Digital Mammography Quality Control Manual were conducted. Image quality evaluations of the patients' mammograms were conducted by the breast radiologists. The DBT procedure data of patients with acceptable image quality evaluations were analysed. The median of the AGD data distribution in this study was determined. Seventy-four study participants, with an age range of 26 to 69 years old and overall compressed breast thickness (CBT) range of 32.25 to 69.5 mm, were utilised in this study. Image qualities of 96% of the study participants were found to be acceptable while the remaining 4% received recommendations to repeat the image acquisition. Tube loading, peak kilovoltage, and CBT were found to significantly affect the AGD for both projections. The overall AGD median was determined to be 2.13 mGy which is within the recommended 3.0 mGy acceptable limit of ACR.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Establishment of local diagnostic reference levels for mammography: optimizing patient radiation dose within Emirates Health ServicesEgyptian Journal of Radiology and Nuclear Medicine · 2026
- Machine learning-based analysis of average glandular dose in mammography: The role of breast density.Radiography (London, England : 1995) · 2026
- Quality Control dan Analisis Korelasi Mean Glandular Dose Terhadap Parameter Fisik pada Pesawat Mamografi Di Instalasi Radiologi RSUP Dr. M. Djamil PadangJurnal Fisika Unand · 2026
- Comparison of Total X-Ray Energy Absorbed by Fibroglandular Tissue and its Mass Derived from Full Field Digital Mammography and Digital Breast Tomosynthesis.bioRxiv : the preprint server for biology · 2026
- Unmasking Malignancy in Dense Breast Tissue (American College of Radiology Types C and D): A Biopsy-Proven Analysis of Digital Breast Tomosynthesis Performance, Clinical Decision Dynamics, and Correlation with Histological Grade and Molecular Subtypes.Diagnostics (Basel, Switzerland) · 2026
- Comparison of two automated PGMI-based software tools for breast positioning quality assessment in DBT screening.European radiology experimental · 2026