Comparison of Total X-Ray Energy Absorbed by Fibroglandular Tissue and its Mass Derived from Full Field Digital Mammography and Digital Breast Tomosynthesis.
John Heine, Erin Fowler, R Jared Weinfurtner, Kathleen M Egan
Key points
From the paper's conclusion- These findings demonstrate proof-of-concept internal consistency but do not establish absolute dosimetric accuracy.
Abstract
This study investigated whether total x-ray energy absorbed by fibroglandular tissue and fibroglandular tissue mass derived from raw and clinical full field digital mammography (FFDM) images corresponded with the respective quantities derived from digital breast tomosynthesis (DBT). A total of 840 paired FFDM and DBT observations were analyzed separately for the craniocaudal (CC) and mediolateral oblique (MLO) orientations. Total absorbed x-ray energy and fibroglandular tissue mass were calculated using an automated breast density detection method in conjunction with the recorded average glandular dose. Relationships between FFDM and DBT quantities were evaluated in the log-log domain using linear and quadratic models, with model performance assessed by R², adjusted R², mean squared error (MSE), chi-square goodness-of-fit tests, and residual analysis. Strong associations were observed between FFDM and respective DBT quantities. For raw FFDM, linear energy models produced R² values of 0.734 and 0.901 for the CC and MLO views, respectively, while the corresponding mass models produced R² values of 0.634 and 0.870. Inclusion of a quadratic term improved model performance modestly. For clinical FFDM, linear energy models produced R² values of 0.854 and 0.919, while the corresponding mass models produced R² values of 0.721 and 0.812. The first-order coefficients for the clinical-image energy models were approximately unity (0.99-1.05), and inclusion of the quadratic term produced negligible changes in model performance and residual distributions. The goodness-of-fit tests did not reject adequate fit for any model. Clinical FFDM images can therefore provide estimates of total absorbed x-ray energy and fibroglandular tissue mass that are strongly associated with corresponding quantities derived from DBT. The approximately unit first-order coefficients and minimal benefit from quadratic modeling support a simple linear log-log calibration for clinical FFDM. This approach may facilitate combining FFDM and DBT based data in larger retrospective and longitudinal studies of cumulative mammographic x-ray exposure. These findings demonstrate proof-of-concept internal consistency but do not establish absolute dosimetric accuracy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Unmasking Malignancy in Dense Breast Tissue (American College of Radiology Types C and D): A Biopsy-Proven Analysis of Digital Breast Tomosynthesis Performance, Clinical Decision Dynamics, and Correlation with Histological Grade and Molecular Subtypes.Diagnostics (Basel, Switzerland) · 2026
- Concordance of Digital Breast Tomosynthesis (DBT) and Magnetic Resonance Imaging (MRI) Measured Tumor Size with the Pathological Gold Standard: A Retrospective Cohort Analysis by Molecular Subtype and Breast Density.Diagnostics (Basel, Switzerland) · 2026
- Machine learning-based analysis of average glandular dose in mammography: The role of breast density.Radiography (London, England : 1995) · 2026
- Fourier Multiscale Analysis of Bilateral Breast Asymmetry Associated with Breast CancermedRxiv · 2026
- Body Size, Radiation Dose and Exposure Index in Conventional Radiography: A Multicentre Prospective Study.Diagnostics (Basel, Switzerland) · 2026
- Radiomic Characterization of Breast Tissue from Breast CT Images Obtained with Synchrotron Beams.Tomography (Ann Arbor, Mich.) · 2026