Systematic identification of human anti-mouse antibody interference causing falsely elevated B-type natriuretic peptide: A case study
Shunchang Li, Jifei Yao
Shaoxing Second Hospital
Key points
From the paper's conclusion- Therefore, we conducted a detailed analysis through a stepwise investigation, including serial dilution (showing non-linear recovery), cross-platform comparison (normal result on an alternative platform), and heterophilic blocking reagent treatment (partial correction).
Abstract
B-type natriuretic peptide (BNP) is crucial for heart failure diagnosis. However, immunoassays are vulnerable to heterophilic antibody interference, which may lead to false-positive results and significant clinical mismanagement. Studies suggest such interference may account for a notable proportion of unexplained biomarker elevations. This report describes a case of a 36-year-old female with a markedly elevated BNP (2735.7 ± 54.3 pg/mL) incongruent with her clinical presentation. Therefore, we conducted a detailed analysis through a stepwise investigation, including serial dilution (showing non-linear recovery), cross-platform comparison (normal result on an alternative platform), and heterophilic blocking reagent treatment (partial correction). Ultimately, the addition of purified mouse IgG, but not anti-goat IgG or IgM blockers, normalized the BNP level in a dose-dependent manner, confirming human anti-mouse antibody (HAMA) interference. This case demonstrates that such interference is one of the potential causes of unexplained elevations in clinical immunoassay results. A systematic workflow was established in the case, ranging from suspicion to confirmation and ultimately to laboratory-adoptable conclusions. This process facilitates accurate diagnosis and helps clinicians avoid unnecessary interventions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- False-Positive hs-cTnI in a Patient on Immunotherapy: a Case of Macromolecular Interference.Clinical laboratory · 2026
- A-046 A novel cardiac troponin assay circumvents positive and negative macrotroponin interferenceClinical Chemistry · 2026
- Frequency of Antibody-Mediated Positive Interference in a Healthy Danish Cohort Using Combined Immunodepletion and Ultracentrifugation with a High-Sensitivity Cardiac Troponin I Assay.The journal of applied laboratory medicine · 2026
- Unexplained Hypercobalaminemia: A Rare Diagnostic Pitfall.EJIFCC · 2026
- A-023 Evaluation of Interfering Substances in VITROS High-Sensitivity Cardiac Troponin I AssayClinical Chemistry · 2026
- Pseudo-Metabolic Alkalosis: A Hidden Laboratory Pitfall With Elevated LDH.The Yale journal of biology and medicine · 2026