Noninterchangeability of Vessel Density Measurements Across Next-Generation SS-OCTA Systems.
Peter M Maloca, Philippe C Cattin, Nadja Inglin, Claudia Bruè, Pascal W Hasler, Simon Rothenbuehler, Hanna Zuche, Javier Zarranz-Ventura,
Department of Biomedical Engineering, University of Basel, Allschwil, Switzerland. · Department of Ophthalmology, University Hospital Basel, Basel, Switzerland.
Key points
From the paper's conclusion- Despite interdevice differences, both devices showed consistent OCTA acquisition.
Abstract
PURPOSE: Comparing swept-source optical coherence tomography angiography (OCTA) vessel density in the superficial and deep capillary plexuses across leading OCTA systems. METHODS: OCTAs were acquired from healthy participants using two OCTA systems (BMizar 400 kHz and Yalkaid 100 kHz, TowardPi, Beijing, China) with scan sizes of 3 × 3, 6 × 6, 12 × 12, and 15 × 15 mm2. The vessel density was quantified using the manufacturer's software. RESULTS: Across 30 subjects comprising 60 eyes and a total of 225 OCTA scans, significant interdevice differences were observed at the Early Treatment Diabetic Retinopathy Study sector level in all eight scan size × plexus comparisons (three-way analysis of variance; P < 1.16 × 10e-4). The mean BMizar vessel density values for the superficial and deep plexuses were 43.4% ± 1.6% and 44.9% ± 1.5%, 44.7% ± 2.7% and 47.9% ± 1.1%, 44.2% ± 1.3% and 45.7% ± 1.1%, and 42.4 ± 1.6 % and 44.3 ± 1.3% for the 3 × 3, 6 × 6, 12 × 12, and 15 × 15 mm² scans; the Yalkaid values were lower. BMizar showed greater vessel density than Yalkaid, particularly in 3 × 3 mm² deep slab scans, with differences of approximately 5% to 10%. Across scans larger than 3 × 3 mm², BMizar showed greater central vessel density; the peripheral patterns varied, with reversals favoring Yalkaid and greater border effects in larger fields of view. CONCLUSIONS: Despite interdevice differences, both devices showed consistent OCTA acquisition. TRANSLATIONAL RELEVANCE: Device differences may affect clinical interpretation, longitudinal follow-up, and multicenter studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantifying segmentation sensitivity in OCTA: Device-specific profiles across three commercial platformsPLoS ONE · 2026
- Comparison of Biometric Measurements Between Anterior Segment Hyperparallel and Swept-Source Optical Coherence Tomography.Translational vision science & technology · 2026
- Reference Percentiles for Macular Vessel Density by OCT Angiography and Their Association with Systemic Diseases in Chinese Adults.Ophthalmology science · 2026
- Within-Session Repeatability of Spectral-Domain OCT Angiography and Inter-Device Agreement with Swept-Source OCT Angiography in Diabetic EyesVision · 2026
- Cross-Device Field-of-View Scale Differences in Clinical OCT Angiography: Phantom Calibration and Large-Scale Clinical Validation.medRxiv : the preprint server for health sciences · 2026
- Mechano-Vascular Considerations in Duane Syndrome: An OCTA Study of Macular Perfusion.Current eye research · 2026