ADAPT: A novel automated discrimination perimetry test for rapid mapping of spared visual discrimination abilities across visual field defects.
Bryan V Redmond, Krystel R Huxlin, Matthew R Cavanaugh
Flaum Eye Institute and Center for Visual Science, University of Rochester Medical Center, Rochester, NY, USA.
Key points
From the paper's conclusion- We validated ADAPT in 18 occipital stroke patients with unilateral, homonymous, visual field deficits, comparing two versions of ADAPT: ADAPTran mimicked HFA test patterns, assessing a new location randomly per trial, while ADAPTseq started in the intact field and then moved sequentially into the blind field.
Abstract
Humphrey Field Analyzers (HFAs) are clinical tools commonly used to diagnose vision loss and measure detection of small, stationary lights. However, laboratory testing has shown that some patients, particularly those with occipital strokes, can detect and discriminate motion inside HFA-defined "blind" fields. No rapid method exists to map the location, quality, and prevalence of these spared visual abilities. To address this gap, we developed an Automated Discrimination Perimetry Test (ADAPT) combining the speed and coverage of the HFA with feature discrimination tasks, performed using eye tracker-enforced fixation. In this first application, ADAPT presented a four-alternative, forced-choice, direction discrimination task with random dot stimuli. We validated ADAPT in 18 occipital stroke patients with unilateral, homonymous, visual field deficits, comparing two versions of ADAPT: ADAPTran mimicked HFA test patterns, assessing a new location randomly per trial, while ADAPTseq started in the intact field and then moved sequentially into the blind field. Patients also underwent HFA perimetry and psychophysical verification (spot testing) of ADAPT-tested locations. Motion discrimination performance was similar for ADAPTran and ADAPTseq and closely approximated spot-testing performance. Blind-field motion abilities were detected in most patients, representing a significant area of vision despite HFA sensitivity far below the criterion for blindness (10 dB). In sum, this report describes a new, automated, fixation-controlled visual test that can be registered to clinical HFA outputs to rapidly screen for preserved visual discrimination abilities across large portions of the visual field. Importantly, the new test's flexible framework allows for easy customization for patients with visual impairments from a range of etiologies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Complementary Visual Assessment: Validation of a High-Precision Eye-Tracking Kinetic Perimetry in the Evaluation of Visual Field Deficits in Hemianopia.Restorative neurology and neuroscience · 2026
- Comparing performance on a continuous visual stimulus tracking task between real and simulated visual field defects: Evidence for long-term adaptation.Journal of vision · 2026
- Where you didn't look: Perimetry based on the analysis of free-viewing gaze behavior while watching movie clips.Journal of vision · 2026
- Natural interaction based visual field prediction in virtual reality - a pilot studymedRxiv · 2026
- Evaluation of the Perimouse Online Suprathreshold Visual Field Test for Glaucoma Screening in a Chinese Population: A Comparison with the Humphrey Field Analyzer.Clinical ophthalmology (Auckland, N.Z.) · 2026
- A Pilot Study to Compare a Portable Tablet Perimeter with Humphrey Field Analyzer in Patients with Neuro-Ophthalmic Disorders.Journal of binocular vision and ocular motility · 2026