Variable Interferences Dictate Different Cutoffs for Aldosterone Measured by Liaison and iSYS Chemiluminescence Immunoassays for the Seated Saline Suppression Test.
Graeme Eisenhofer, Andrea R Horvath, Jun Yang, Carmina T Fuss, Hanna Remde, Martin Reincke, Tracy Ann Williams, Felix Beuschlein,
Department of Medicine III University Hospital Carl Gustav Carus, Technische Universität Dresden Dresden Germany. · New South Wales Health Pathology, Department of Chemical Pathology Prince of Wales Hospital and School Public Health, University of Sydney Sydney NSW Australia.
Key points
From the paper's conclusion- Cutoffs for CLIA-based measurements of aldosterone during the seated saline suppression test are higher than those for LC-MS/MS and vary according to the specific immunoassay. Gray areas around cutoffs, particularly extensive for the Liaison CLIA, must also be considered.
Abstract
BACKGROUND: Variable macromolecular-related interferences with different chemiluminescence immunoassays (CLIA) of plasma aldosterone, particularly at low concentrations during confirmatory tests, are a previously unrecognized problem. We therefore hypothesized that cutoffs for measurements of aldosterone for the seated saline suppression test require tailoring according to the particular immunoassay. METHODS: Multicenter study involving 609 patients with suspected primary aldosteronism who underwent seated saline suppression tests with aldosterone measured by liquid chromatography-mass spectrometry (LC-MS/MS) compared with iSYS or Liaison CLIAs in 285 and 286 respective patients and a different LC-MS/MS method in 38 others. Cutoffs for plasma aldosterone were derived from receiver operating characteristic curves after classification of patients with (n=181) and without primary aldosteronism (n=336) by methods independent of confirmatory tests. For 92 patients classifications were not achieved. RESULTS: Youden's-index derived cutoffs for postsaline infusion plasma aldosterone were 300 and 272 pmol/L for measurements by respective Liaison and iSYS CLIAs compared with 169 pmol/L by LC-MS/MS. At those cutoffs, respective diagnostic sensitivities for Liaison and iSYS immunoassays and LC-MS/MS were 90% (CI, 84-97), 91% (CI, 86-97), and 98% (CI, 96-100) with specificities of 86% (CI, 80-91), 94% (CI, 80-91), and 90% (CI, 86-93). Improved disease confirmation according to specificities of 95% for measurements by Liaison and iSYS CLIAs and LC-MS/MS were provided by respective cutoffs of 436, 297, and 243 pmol/L but with reduced sensitivities of 64% (CI, 53-75), 85% (CI, 78-92), and 78% (CI, 72-84). CONCLUSIONS: Cutoffs for CLIA-based measurements of aldosterone during the seated saline suppression test are higher than those for LC-MS/MS and vary according to the specific immunoassay. Gray areas around cutoffs, particularly extensive for the Liaison CLIA, must also be considered. REGISTRATION: https://drks.de/search/en/trial/DRKS00017084.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Urinary Aldosterone Analysed by LC-MS/MS vs Immunoassay for Diagnosis of Primary Aldosteronism.The Journal of clinical endocrinology and metabolism · 2026
- A diagnostic strategy for primary aldosteronism diagnosis based on urinary aldosterone assay.Clinical chemistry and laboratory medicine · 2026
- Seated Saline Suppression Test for Lateralizing Primary AldosteronismHypertension · 2026
- Integration of 24-h Urine Aldosterone & Tetrahydroaldosterone With Plasma Screening for Primary Aldosteronism: A 7-Year Study.Journal of clinical hypertension (Greenwich, Conn.) · 2026
- A-154 Comparison of Aldosterone: Renin Ratio by DiaSorin immunoassay vs LCMS for Primary Aldosteronism screeningClinical Chemistry · 2026
- Application of the latest CLSI EP guidelines in light-initiated chemiluminescent assay for testosterone.Journal of medical biochemistry · 2026