Variable Interferences Dictate Different Cutoffs for Aldosterone Measured by Liaison and iSYS Chemiluminescence Immunoassays for the Seated Saline Suppression Test.

Graeme Eisenhofer, Andrea R Horvath, Jun Yang, Carmina T Fuss, Hanna Remde, Martin Reincke, Tracy Ann Williams, Felix Beuschlein,

Department of Medicine III University Hospital Carl Gustav Carus, Technische Universität Dresden Dresden Germany. · New South Wales Health Pathology, Department of Chemical Pathology Prince of Wales Hospital and School Public Health, University of Sydney Sydney NSW Australia.

Key points

From the paper's conclusion
  • Cutoffs for CLIA-based measurements of aldosterone during the seated saline suppression test are higher than those for LC-MS/MS and vary according to the specific immunoassay. Gray areas around cutoffs, particularly extensive for the Liaison CLIA, must also be considered.

Abstract

BACKGROUND: Variable macromolecular-related interferences with different chemiluminescence immunoassays (CLIA) of plasma aldosterone, particularly at low concentrations during confirmatory tests, are a previously unrecognized problem. We therefore hypothesized that cutoffs for measurements of aldosterone for the seated saline suppression test require tailoring according to the particular immunoassay. METHODS: Multicenter study involving 609 patients with suspected primary aldosteronism who underwent seated saline suppression tests with aldosterone measured by liquid chromatography-mass spectrometry (LC-MS/MS) compared with iSYS or Liaison CLIAs in 285 and 286 respective patients and a different LC-MS/MS method in 38 others. Cutoffs for plasma aldosterone were derived from receiver operating characteristic curves after classification of patients with (n=181) and without primary aldosteronism (n=336) by methods independent of confirmatory tests. For 92 patients classifications were not achieved. RESULTS: Youden's-index derived cutoffs for postsaline infusion plasma aldosterone were 300 and 272 pmol/L for measurements by respective Liaison and iSYS CLIAs compared with 169 pmol/L by LC-MS/MS. At those cutoffs, respective diagnostic sensitivities for Liaison and iSYS immunoassays and LC-MS/MS were 90% (CI, 84-97), 91% (CI, 86-97), and 98% (CI, 96-100) with specificities of 86% (CI, 80-91), 94% (CI, 80-91), and 90% (CI, 86-93). Improved disease confirmation according to specificities of 95% for measurements by Liaison and iSYS CLIAs and LC-MS/MS were provided by respective cutoffs of 436, 297, and 243 pmol/L but with reduced sensitivities of 64% (CI, 53-75), 85% (CI, 78-92), and 78% (CI, 72-84). CONCLUSIONS: Cutoffs for CLIA-based measurements of aldosterone during the seated saline suppression test are higher than those for LC-MS/MS and vary according to the specific immunoassay. Gray areas around cutoffs, particularly extensive for the Liaison CLIA, must also be considered. REGISTRATION: https://drks.de/search/en/trial/DRKS00017084.

Read the paper