Midterm Outcomes of Intra-Annular Versus Supra-Annular Self-Expanding Transcatheter Aortic Valves.
Takuya Ogami, Francis Ferdinand, Xander Jacquemyn, Irsa Hasan, Yisi Wang, Derek Serna-Gallegos, Dustin Kliner, Amber Makani,
Division of Cardiac Surgery, Department of Cardiothoracic Surgery University of Pittsburgh Pittsburgh PA USA. · Division of Cardiology University of Pittsburgh Pittsburgh PA USA.
Key points
From the paper's conclusion- Permanent pacemaker placement was significantly higher in the intra-annular group. This may be related to only high-risk approval for intra-annular valves. Survival and other complications were comparable between the groups.
Abstract
BACKGROUND: Intra-annular self-expanding valves have been a relatively recent addition to transcatheter heart valves in the United States when compared with the more established supra-annular self-expanding valve. Although many studies have compared outcomes between self-expanding and balloon-expandable valves, evidence specifically comparing intra-annular and supra-annular self-expanding valves is sparse. METHODS: This is a retrospective observational study including all patients who underwent transcatheter aortic valve replacement with a self-expanding valve in our institutional TVT (Transcatheter Valve Therapy) registry between 2014 and 2024. Patients were grouped between the intra-annular and supra-annular groups. RESULTS: Of 5470 patients undergoing transcatheter aortic valve replacement in the time period, a total of 2294 patients underwent transcatheter aortic valve replacement with self-expanding valves, including 356 (18.4%) with intra-annular and 1938 (84.5%) with supra-annular. Propensity-score matching provided 354 patients in each group. After propensity-score matching, the incidence of new permanent pacemaker placement was higher in the intra-annular group (11.9% versus 3.4% in the supra-annular group, P<0.001). Moderate or greater paravalvular regurgitation was only seen in 1.6% (n=11) overall. Although mild paravalvular regurgitation was more frequently seen in the intra-annular group at the immediate postoperative echocardiogram (22.0% versus 11.9% in the supra-annular group, P<0.001), its statistical differences diminished at 1 year (5.1% in the intra-annular group versus 8.5% in the supra-annular group, P=0.07). The midterm survival was comparable between groups at 2 years. A Cox hazard analysis demonstrated that the type of valve was not associated with all-cause mortality (hazard ratio, 1.06 [95% CI, 0.72-1.57], P=0.77). CONCLUSIONS: Permanent pacemaker placement was significantly higher in the intra-annular group. This may be related to only high-risk approval for intra-annular valves. Survival and other complications were comparable between the groups.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Outcomes of Supra-Annular vs Intra-Annular Self-Expanding Valves in Valve-In-Valve Transcatheter Aortic Valve Replacement.JACC. Cardiovascular interventions · 2026
- Clinical Outcomes and Valve Performance of Intra-Annular Self-Expanding Transcatheter Aortic Valve in Valve-in-Valve Procedures: Insights From the NAViV Registry.Journal of the American Heart Association · 2026
- Five-Year Follow-Up of Transcatheter Aortic Valve Replacement With Self-Expanding Intra-Annular Versus Commercially Available Valves.JACC. Cardiovascular interventions · 2026
- Long-term outcomes after balloon- or self-expandable transcatheter aortic valve replacement in small aortic annuli.Cardiovascular intervention and therapeutics · 2026
- Comparative outcomes of balloon-expandable versus self-expanding transcatheter aortic valves: insights from a large-scale German registry.BMC cardiovascular disorders · 2026
- Balloon-Expandable Vs Self-Expanding Valves for Transcatheter Aortic Valve Implantation in Internally Stented Surgical Valves.CJC open · 2026