Effects of Vibrating Mesh Versus Jet Nebulization on Ventilatory Pressures During Non-Invasive Ventilation: An In Vitro Study.

David Wood, Katherine Stöwhas, Daniela Aguilar, Camila Aguilera, Carlos Jorquera, Guillermo Droppelmann

Faculty of Medicine, School of Kinesiology, Universidad Finis Terrae, Santiago 7501015, Chile. · Faculty of Sciences, School of Nutrition and Dietetics, Universidad Mayor, Santiago 8580745, Chile.

Preclinical

Key points

From the paper's conclusion
  • These findings suggest that nebulizer technology may influence pressure behavior during NIV under the experimental conditions evaluated.

Abstract

The integration of aerosol therapy with non-invasive ventilation (NIV) is a widespread clinical practice, although its effects on key ventilatory pressures remain insufficiently characterized. Vibrating mesh nebulizers (VMNs) provide superior drug delivery efficiency; however, their influence on inspiratory and expiratory positive airway pressures (IPAP and EPAP) compared to jet nebulizers (JNs) is not well defined. This study aimed to evaluate the effects of VMNs and JNs on IPAP and EPAP during NIV using a controlled in vitro model. An in vitro experimental study was performed employing a 1 L test lung coupled to a BIPAP Synchrony ventilator. Three experimental conditions were investigated: control (no nebulization), VMN (Aerogen® Solo), and JN (Hudson RCI® Micro Mist®). A total of 11 repeated measurements were obtained per condition over a 5 min period, during which proximal and distal pressures were continuously recorded. Data were analyzed via two-way repeated-measure ANOVA, with Cohen's d employed to quantify effect sizes. Jet nebulization was associated with significant increases in EPAP at both proximal (8.12 ± 0.1 vs. 7.9 ± 0.0 cmH2O; p < 0.001) and distal measurement sites (7.7 ± 0.1 vs. 7.5 ± 0.0 cmH2O; p < 0.001), exhibiting very large effect sizes (d = 3.95-4.80). In contrast, VMN maintained EPAP at levels comparable to those of the control (d = 0.25, p = 0.96). IPAP values remained stable across all tested conditions, with no statistically significant differences observed. In this in vitro model, jet nebulizers were associated with higher expiratory airway pressures during NIV, whereas vibrating mesh nebulizers preserved pressure values comparable to those of control conditions. These findings suggest that nebulizer technology may influence pressure behavior during NIV under the experimental conditions evaluated. However, the absolute pressure differences observed were small (approximately 0.23-0.24 cmH2O) and are unlikely to be clinically relevant. Further investigation in clinical settings is warranted to determine whether these mechanical differences translate into clinically meaningful outcomes.

Read the paper

Discussion