Effects of Vibrating Mesh Versus Jet Nebulization on Ventilatory Pressures During Non-Invasive Ventilation: An In Vitro Study.
David Wood, Katherine Stöwhas, Daniela Aguilar, Camila Aguilera, Carlos Jorquera, Guillermo Droppelmann
Faculty of Medicine, School of Kinesiology, Universidad Finis Terrae, Santiago 7501015, Chile. · Faculty of Sciences, School of Nutrition and Dietetics, Universidad Mayor, Santiago 8580745, Chile.
Key points
From the paper's conclusion- These findings suggest that nebulizer technology may influence pressure behavior during NIV under the experimental conditions evaluated.
Abstract
The integration of aerosol therapy with non-invasive ventilation (NIV) is a widespread clinical practice, although its effects on key ventilatory pressures remain insufficiently characterized. Vibrating mesh nebulizers (VMNs) provide superior drug delivery efficiency; however, their influence on inspiratory and expiratory positive airway pressures (IPAP and EPAP) compared to jet nebulizers (JNs) is not well defined. This study aimed to evaluate the effects of VMNs and JNs on IPAP and EPAP during NIV using a controlled in vitro model. An in vitro experimental study was performed employing a 1 L test lung coupled to a BIPAP Synchrony ventilator. Three experimental conditions were investigated: control (no nebulization), VMN (Aerogen® Solo), and JN (Hudson RCI® Micro Mist®). A total of 11 repeated measurements were obtained per condition over a 5 min period, during which proximal and distal pressures were continuously recorded. Data were analyzed via two-way repeated-measure ANOVA, with Cohen's d employed to quantify effect sizes. Jet nebulization was associated with significant increases in EPAP at both proximal (8.12 ± 0.1 vs. 7.9 ± 0.0 cmH2O; p < 0.001) and distal measurement sites (7.7 ± 0.1 vs. 7.5 ± 0.0 cmH2O; p < 0.001), exhibiting very large effect sizes (d = 3.95-4.80). In contrast, VMN maintained EPAP at levels comparable to those of the control (d = 0.25, p = 0.96). IPAP values remained stable across all tested conditions, with no statistically significant differences observed. In this in vitro model, jet nebulizers were associated with higher expiratory airway pressures during NIV, whereas vibrating mesh nebulizers preserved pressure values comparable to those of control conditions. These findings suggest that nebulizer technology may influence pressure behavior during NIV under the experimental conditions evaluated. However, the absolute pressure differences observed were small (approximately 0.23-0.24 cmH2O) and are unlikely to be clinically relevant. Further investigation in clinical settings is warranted to determine whether these mechanical differences translate into clinically meaningful outcomes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of aerosol delivery in ventilated patients via vibrating mesh nebuliser versus other delivery devices using a patient-specific in silico whole-lung model.Respiratory research · 2026
- Vibrating mesh nebuliser positioning during helmet CPAP: A thermo-hygrometric bench-top study.Intensive & critical care nursing · 2026
- Bronchodilator Drug Delivery With Pressurized Meter-Dose-Inhaler Versus Jet Nebulizer With Noninvasive-Ventilation in Acute-Exacerbation of COPD: A Randomized Controlled Pharmacokinetic Endpoint Trial.Respirology (Carlton, Vic.) · 2026
- Effect of Nebulizer Breath-Guiding on Breathing Patterns and Lung Deposition.Journal of aerosol medicine and pulmonary drug delivery · 2026
- Design and evaluation of a breath-synchronized smart vibrating mesh nebulizer: from instrumentation to human pharmacokinetics.International journal of pharmaceutics · 2026
- The Effect of Independent Humidity Control on Aerosol Drug Delivery During Adult Invasive Mechanical VentilationRespiratory Care · 2026