  • Dentinal defects were detected at different root levels following preparation with all three rotary systems, with higher overall defect incidence than in the unprepared control group; however, the present methodology does not allow definitive attribution of these defects to instrumentation.

Abstract

Endodontic treatment relies on effective root canal preparation using nickel-titanium (NiTi) rotary instruments; however, dentinal defects may be observed following mechanical canal preparation. A total of 40 extracted human mandibular premolars with single roots and single straight canals were randomly allocated into four groups (n = 10), control (unprepared), TruNatomy (TRN), ProTaper Next (PTN), and Dia-X. Root canal was prepared using an ENDO-MATE DT endodontic motor (NSK, Japan), and horizontally sectioned from the apex using a low-speed diamond disc, corresponding to apical, middle, and coronal root thirds, respectively. Dentinal microcracks were evaluated under a stereomicroscope at ×20 magnification. Dentinal defects were observed in all instrumented groups, while the unprepared control group showed the lowest incidence, with defects detected at the apical and coronal thirds (10% each). No statistically significant differences were found among the three rotary systems at the apical (p = 0.211), middle (p = 0.069), or coronal (p = 0.600) levels. However, overall comparison showed a statistically significant difference among groups (p = 0.043), with significant differences between the control group and TRN (p = 0.045), PTN (p = 0.045), and Dia-X (p = 0.010). Dentinal defects were detected at different root levels following preparation with all three rotary systems, with higher overall defect incidence than in the unprepared control group; however, the present methodology does not allow definitive attribution of these defects to instrumentation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Comparative Assessment of Observed Dentinal Defects Following Root Canal Preparation With TruNatomy, ProTaper Next, and Dia-X Nickel-Titanium Rotary Systems: An In Vitro Study — Science Skim