Fluoride-Oxalate and Lithium Heparin Anticoagulant Tubes Are Reliable Alternatives to EDTA for HbA1c Measurement.
Lamia Abdellaoui, Asma Krir, Manel Soltani, Aouatef Trabelsi, Mehdi Mrad, Afef Bahlous
Department of Clinical Biochemistry, Pasteur Institute of Tunis, University of Tunis El Manar, Tunis, Tunisia, pasteur.tn.
Key points
From the paper's conclusion- Fluoride-oxalate and lithium heparin anticoagulant tubes provide reliable alternatives to standard K3-EDTA tubes for HbA1c measurement, with no clinically meaningful differences observed. However, we emphasize that these results are strictly validated for the Roche COBAS immunoassay.
Abstract
BACKGROUND: The dual role of HbA1c in diagnosing and monitoring diabetes mellitus has driven a continuous increase in routine testing requests. To optimize laboratory workflows and avoid dedicated EDTA blood draws, this study is aimed at determining whether fluoride-oxalate and lithium heparin anticoagulant tubes could serve as reliable alternatives to EDTA tubes for HbA1c measurement on the COBAS INTEGRA 400 plus analyzer. METHODS: Over 2 months at the Pasteur Institute of Tunis, paired blood samples from patients undergoing routine HbA1c testing were analyzed. Standard K3-EDTA tubes were compared with either lithium heparin or fluoride-oxalate tubes using a turbidimetric inhibition immunoassay on a COBAS INTEGRA 400 plus analyzer. The study relied exclusively on residual clinical material without imposing specific preanalytical restrictions. RESULTS: The study population comprised 277 patients, divided into Group 1 (K3-EDTA vs. fluoride-oxalate, n = 200) and Group 2 (K3-EDTA vs. lithium heparin, n = 77). HbA1c results from both alternative tubes showed strong correlation with K3-EDTA (r = 0.99, p < 0.0001), with no significant overall differences. Passing-Bablok regression revealed no constant or proportional bias for lithium heparin, whereas a systematic bias was observed for fluoride-oxalate. Bland-Altman analysis of log-transformed data indicated no significant mean bias for either fluoride-oxalate (-0.15%; 95% LoA [-2.76% to +2.53%]) or lithium heparin (-0.04%; 95% LoA [-2.77% to +2.76%]), although a slight proportional bias was detected in both. Finally, HbA1c measurements in both tube types exhibited high diagnostic performance at clinical cutoffs (p < 0.0001). CONCLUSIONS: Fluoride-oxalate and lithium heparin anticoagulant tubes provide reliable alternatives to standard K3-EDTA tubes for HbA1c measurement, with no clinically meaningful differences observed. However, we emphasize that these results are strictly validated for the Roche COBAS immunoassay.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative analysis of enzymatic assay for HbA1c measurement with high performance liquid chromatography.Scientific reports · 2026
- Method comparison and agreement of a fluorescence immunoassay-based HbA1c test with a routine HPLC method.Diabetology international · 2026
- Evaluation of HbA1c stability in refrigerated blood samples: for how long are samples fit for use?Scandinavian journal of clinical and laboratory investigation · 2026
- Evaluation of the Lifotronic H100 Plus analyzer for HbA1c determination.Scandinavian journal of clinical and laboratory investigation · 2026
- B-346 Storage Stability of HbA1c in Whole Blood: Method-Dependent Interference on Ion-Exchange HPLC and Degradation Kinetics Characterized by LC–MSClinical Chemistry · 2026
- B-005 Analytical Performance Evaluation of Two HbA1c Testing Methods on Roche cobas c513 AnalyzersClinical Chemistry · 2026