Incidence and risk factors for delayed structural changes in Pipeline Embolization Devices
Clint Badger, Souvik Singha, Huanwen Chen, Matthew Arnold, Najme Hosseini, Rosy Linda Njonkou-Tchoquessi, Mohamed A. Labib, Shyam Majmundar,
University of Maryland, Baltimore
Key points
From the paper's conclusion- We retrospectively analyzed all Pipeline Embolization Device (PED) treatments for intracranial aneurysms at our institution from November 2011 to April 2024.
Abstract
BACKGROUND: Flow-diverting stents have transformed the management of intracranial aneurysms, but delayed structural changes of these devices, or delayed braid deformation (DBD), may compromise the parent vessel and increase thromboembolic risk. Data on the incidence, evolution, and predictors of DBD remain limited. METHODS: We retrospectively analyzed all Pipeline Embolization Device (PED) treatments for intracranial aneurysms at our institution from November 2011 to April 2024. Patients without follow-up imaging or treatment with other flow diverters were excluded. Demographic, aneurysm, and procedural variables were collected, including age, vessel-device size mismatch, and intraprocedural vasodilator use. Follow-up imaging was reviewed for DBD, parent-vessel narrowing, and angiographic or clinical sequelae. Multivariable logistic regression identified independent predictors. RESULTS: Among 566 PED treatments, DBD occurred in 59 (10.4%). Most cases were mild (≤49% luminal narrowing), but six (10.2%) showed ≥50% compromise, and two required balloon angioplasty for symptomatic flow limitation. Independent predictors included younger age (OR 1.51 per 10 year decrease, P<0.001) and device oversizing ≥1 mm (OR 3.14, P<0.001). The use of intraprocedural verapamil was not significant but showed a strong statistical trend (OR 2.31, P=0.053). Hypertension, aneurysm location, and PED generation were not associated with DBD. Among patients with imaging follow-up after diagnosis, >50% of non-treated cases showed spontaneous improvement or resolution. CONCLUSIONS: DBD of the PED after aneurysm treatment is relatively common but usually mild and self-limited. Younger age and device oversizing independently increase the risk of the phenomenon.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Incidence and Predictors of Braid Deformations following Aneurysm Treatment with the DERIVO(2heal) Flow Diverter: A Single-Center Experience.AJNR. American journal of neuroradiology · 2026
- Association between the Stent-to-Distal Parent Artery Diameter Ratio and Diffusion-Weighted Imaging-Positive Lesion Burden after Pipeline Shield Treatment.Journal of neuroradiology = Journal de neuroradiologie · 2026
- First clinical experience with the surpass elite flow diverter in the treatment of intracranial aneurysmsNeuroradiology · 2026
- A Multicenter Retrospective Study on the Safety and Efficacy of the Surpass Evolve Flow Diverter System for the Treatment of Intracranial AneurysmsmedRxiv · 2026
- Delayed Aneurysmal Rupture following Flow Diverter Placement despite Adjunctive Coiling: Case Report.Case reports in neurology · 2026
- Flow diversion versus conventional endovascular treatment for ruptured basilar artery dissecting aneurysms: a single-center cohort study.Frontiers in neurology · 2026