Clinical performance and practical implications of continuous glucose monitoring in very preterm infants: a cohort study.
Gordon Xin Hua Liu, Loredana Marcovecchio, Kathryn Beardsall, REACT Collaborative
School of Medicine, University of Auckland, Grafton Campus, Auckland, New Zealand. · Department of Paediatrics, University of Cambridge, Cambridge, UK.
Key points
From the paper's conclusion- Objective To explore the clinical performance of continuous glucose monitoring (CGM) in very preterm infants during the first week of life, using data from the multicentre Real time continuous glucose monitoring (REACT) randomised controlled trial (RCT) and to consider how CGM can best support neonatal glucose management.
Abstract
OBJECTIVE: To explore the clinical performance of continuous glucose monitoring (CGM) in very preterm infants during the first week of life, using data from the multicentre Real time continuous glucose monitoring (REACT) randomised controlled trial (RCT) and to consider how CGM can best support neonatal glucose management. DESIGN: Prespecified secondary analysis of paired sensor glucose (SG) and blood glucose (BG) measurements from the REACT RCT. SETTING: 13 neonatal intensive care units (NICUs) in the UK, Spain and the Netherlands. PATIENTS: 155 very preterm infants (≤1200 g birth weight or <34 weeks' gestation) with CGM data collected during the first week of life. MAIN OUTCOME MEASURES: Mean absolute relative difference (MARD), ISO 15197:2013 system agreement, Bland-Altman bias, error grid analysis and clinical contextualisation of CGM performance. RESULTS: Across 2223 SG-BG pairs, overall MARD was 11.9%, and 98%-99% of values fell within clinically acceptable error grid zones. SG tended to be less accurate at BG extremes, although error grid analyses show that clinical risk from discrepancies was low. CGM detected episodes of dysglycaemia not captured by intermittent sampling. Glucose variability did not materially influence accuracy. CONCLUSIONS: CGM can provide clinically meaningful information in very preterm infants despite modest analytical inaccuracy when compared with single point-of-care values. Its greatest utility lies in continuous trend monitoring, early detection of silent dysglycaemia and reduced invasive sampling, rather than replacement of diagnostic blood testing. Clinicians should interpret SG values in context, confirm unexpected extremes with BG and focus on patterns rather than single values.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The neonatal continuous glucose monitoring paradox: why at-risk infants still lack access to a transformative technology-A call for needs-led innovation.Frontiers in medicine · 2026
- Comparing efficacy in glycaemic control and pregnancy outcomes between continuous glucose monitoring and self-monitoring of blood glucose in pregnancies complicated by gestational diabetes mellitus: a systematic review and meta-analysis.Diabetes research and clinical practice · 2026
- Glycemic variability and reference percentiles in very low birth weight preterm infants using continuous glucose monitoringPLoS ONE · 2026
- Performance and Safety of a Continuous Glucose Monitoring System in Pregnant Women With Gestational Diabetes Mellitus: A Multicenter, Open-Label TrialJournal of Diabetes Science and Technology · 2026
- Continuous Glucose Monitoring in Gestational Diabetes Mellitus: Evidence, Clinical Benefits and Remaining Challenges.Diabetes therapy : research, treatment and education of diabetes and related disorders · 2026
- Clinical accuracy of Dexcom G6 and G7 continuous glucose monitors in hospitalized pediatric patients with type 1 diabetes: a real-world studyAnnals of Pediatric Endocrinology & Metabolism · 2026