Serotonin measurement by lab-on-a-contact lens in porcine and human tears.

Minwoo Kim, Sang Won Lee, Chenguang Zhang, Wenyi Zeng, Liangshu Liu, Aliesha Danielle Garrett, Reihaneh Haghniaz, Dawn Meyer,

Biomedical Device Center, Terasaki Institute for Biomedical Innovation, Los Angeles, CA 91367, USA. · Department of Biomedical Engineering, Yonsei University, Wonju 26493, Republic of Korea.

Trial

Key points

From the paper's conclusion
  • Here, we present a reversible lab-on-a-contact lens platform that integrates a printable, stretchable ferrocene-based electrochemical sensor for in situ detection of serotonin in tear fluid.

Abstract

Stress-related endocrine responses are critical for understanding human performance and mental health. Conventional assessments rely heavily on subjective self-report questionnaires or invasive sampling such as venipuncture. Tears offer a noninvasive source of neurochemical biomarkers, including serotonin, a key regulator of stress and mood. However, real-time and continuous monitoring of tear serotonin has not yet been fully realized. Here, we present a reversible lab-on-a-contact lens platform that integrates a printable, stretchable ferrocene-based electrochemical sensor for in situ detection of serotonin in tear fluid. The device is fabricated on a soft hydrogel lens and uses an electrochemically refreshable redox surface to restore baseline between measurements. Calibration in artificial tears established a limit of detection of 72 picomolar and a dynamic range of 0.1 to 500 nanomolar. Function was maintained for more than 28 days in a tear-like environment and under repeated mechanical deformations (flipping, folding, stretching, and twisting), with stability confirmed by periodic redox cycling. Ocular safety and biocompatibility were demonstrated in a porcine eye model. In a chronic stress mouse model, tear serotonin concentrations inversely correlated with cortisol in both serum and tears, and a first-in-human pilot study further supported the relationship between tear serotonin and stress with off-line analysis by the lab-on-a-contact lens platform. This platform could enable applications in decentralized, noninvasive endocrine monitoring and holds promise as a clinically translatable tool for real-time mental health biomarker assessment.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Serotonin measurement by lab-on-a-contact lens in porcine and human tears — Science Skim