Validity and Reliability of the 3-Dimensional Printed PhysioCode Handheld Dynamometer for Knee Muscle Strength Assessment.
George Schayer Sabino, Diego Henrique Antunes Nascimento, Fernanda Oliveira Madaleno, Claysson Bruno Santos Vimieiro, Pollyana Helena Vieira Costa, Guilherme Augusto Santos Araujo, Matheus Hissa Lourenço Ferreira, Renan Alves Resende
Department of Physical Therapy, Graduate Program in Rehabilitation Sciences, School of Physical Education, Physical Therapy and Occupational Therapy, Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG, Brazil. · Bioengineering Laboratory, Graduate Program in Mechanical Engineering, Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG, Brazil.
Key points
From the paper's conclusion- The 3D-printed PhysioCode showed adequate concurrent validity and excellent test-retest reliability for knee extensor/flexor strength, with performance comparable to the MicroFET II in this sample of young, healthy adults. Averaging 2 or 3 repetitions improved precision. These findings suggest that PhysioCode may represent an option for strength assessment, warranting further evaluation in target clinical populations.
Abstract
CONTEXT: Handheld dynamometry is common for isometric strength testing, but the high cost and import barriers of commercial devices (eg, MicroFET II) limit access. Three-dimensional (3D) printing enables low-cost, customizable instruments. We developed the 3D-printed PhysioCode and examined its measurement properties for knee strength. DESIGN: Methods and Technical Report evaluating concurrent validity and test-retest reliability versus a reference handheld dynamometer. METHODS: Forty-nine healthy adults (23.2 [3.9] y; body mass index 21.9 [2.8] kg/m2) performed 3 isometric trials of dominant-limb knee extension and flexion with both devices in randomized order, assessed by 2 trained novice raters under blinded analysis. Concurrent validity was evaluated using Pearson correlation coefficients (r) and root mean square error, and agreement between devices was assessed using Bland-Altman analyses. Reliability was assessed using intraclass correlation coefficients (ICC[2,1] and ICC[2,k]), along with the standard error of measurement and minimal detectable change at 95% confidence. We compared single and repeated test strategies. RESULTS: All mean values differed by ≤21 N between devices across knee extensors and flexors. Concurrent validity was strong-very strong (r = .88-.95; P < .001), with lower root mean square error for averaged repetitions. Bland-Altman showed small, direction-specific bias: PhysioCode slightly overestimated extensor force values (-17 to 21 N) and underestimated flexor force values (-12 N). Reliability was excellent: PhysioCode ICC(2,1) = .93 to .95 and ICC(2,k) = .98 to .98; MicroFET II ICC(2,1) = .89 to .94 and ICC(2,k) = .96 to .98. Averaging 2 to 3 trials reduced standard error of measurement and minimal detectable change at 95% confidence relative to single trials or maxima. CONCLUSIONS: The 3D-printed PhysioCode showed adequate concurrent validity and excellent test-retest reliability for knee extensor/flexor strength, with performance comparable to the MicroFET II in this sample of young, healthy adults. Averaging 2 or 3 repetitions improved precision. These findings suggest that PhysioCode may represent an option for strength assessment, warranting further evaluation in target clinical populations.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development, Calibration and Metrological Characterization of a Low-Cost 3D Printed Handgrip Dynamometer Based on Force-Sensing Resistor TechnologyMetrology · 2026
- A Novel Method of Clinical Strength Assessment: Reliability and Validity of a Crane Scale for the Quadriceps and HamstringsThe Journal of Strength and Conditioning Research · 2026
- Portable Dynamometry Is Reliable for Measuring Knee and Ankle Strength in Daily Practice: Results From a Systematic Review With Meta-analysis and Evidence Gap Map Using the COSMIN Methodology.The Journal of orthopaedic and sports physical therapy · 2026
- Measurement Properties of Hand‐Held Dynamometry for Ankle Strength Assessment: A COSMIN‐Based Systematic Review and Meta‐AnalysisMusculoskeletal Care · 2026
- Agreement Between Handheld and Fixed Dynamometry for Hip Strength Assessment in Young Soccer Players and Non-Soccer Controls: An Exploratory Study.Journal of clinical medicine · 2026
- Digital handgrip assessment in musculoskeletal health: reliability, measurement precision, interchangeability, and clinical implications for rehabilitation and sports medicineBMC Sports Science, Medicine and Rehabilitation · 2026