Key points
From the paper's conclusion- Spatial resolution improved with a threshold number up to about 8 to 10, beyond which additional programmable thresholds gave only modest gain. The achieved tangential resolution also suggests that analyzer-free phase-contrast imaging is feasible for realistic geometries and expected interference patterns.
Abstract
PURPOSE: We are developing a monolithic deep-silicon photon-counting sensor targeting spatial resolution on the order of 1 μ m . This work investigates how pixel pitch, noise level, threshold number, and threshold placement affect the achievable tangential and wafer-thickness resolution to guide the CMOS electronics and sensor design. APPROACH: Allpix Squared simulations were used to evaluate pixel pitches with two noise levels and two threshold-placement schemes over varying threshold numbers. Interaction position was estimated in both dimensions using lookup tables trained on the thresholded pixel output, and performance was quantified using the MTF. RESULTS: Decreasing pixel pitch strongly improved tangential resolution for Compton interactions, whereas photoelectric interactions showed a weaker dependence on pixel pitch and threshold placement. Equal-counting thresholds approached the achievable resolution with fewer thresholds than equidistant thresholds. For eight equal-counting thresholds, the 10% MTF across the studied pixel sizes ranged from 1284 lp / cm to 2452 lp / cm in the tangential direction and from 15 to 71 lp / cm in the wafer-thickness direction for Compton interactions, compared with 409 to 507 lp / cm and 18 to 79 lp / cm , respectively, for photoelectric interactions. RMSE analysis further showed that some interactions for the 25 μ m pitch reached below 2 μ m tangentially and a few tens of micrometers in the wafer-thickness direction. CONCLUSIONS: Spatial resolution improved with a threshold number up to about 8 to 10, beyond which additional programmable thresholds gave only modest gain. The achieved tangential resolution also suggests that analyzer-free phase-contrast imaging is feasible for realistic geometries and expected interference patterns.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Single-Acquisition Spectral and Ultra-High-Resolution Pediatric Head CT: Phantom Comparison of Deep-Silicon Photon-Counting and Energy-Integrating DetectorsmedRxiv · 2026
- Phantom evaluation of small-pixel effect in ultra-high-resolution photon-counting CT: Noise texture and high-contrast spatial resolution.Journal of applied clinical medical physics · 2026
- Small-pixel Acquisition Unmasks "Bucket Effects" of Iterative Reconstruction in Ultra-High-Resolution Photon-Counting CT.Academic radiology · 2026
- Spatial resolution for gamma ray interactions in stacks of monolithic silicon sensorsJournal of Instrumentation · 2026
- Acquisition Speed versus Spatial Resolution in Ultra-High-Resolution Photon-Counting CT: Phantom Study to Guide Protocol DevelopmentmedRxiv · 2026
- Efficiently harvesting low‐energy photons from silicon photon counting detectorsMedical Physics · 2026