Phantom evaluation of small-pixel effect in ultra-high-resolution photon-counting CT: Noise texture and high-contrast spatial resolution.
Kyu-Ho Song, Colin Shan, Gary Xu, Xinhui Duan, Yue Zhang, Kuan Zhang, Liqiang Ren
Department of Radiology, UT Southwestern Medical Center, Dallas, Texas, USA. · Imaging Services, UT Southwestern Medical Center, Dallas, Texas, USA.
Key points
From the paper's conclusion- The UHR mode of PCCT consistently achieves lower image noise than the STD mode without compromising high-contrast spatial resolution. Its greatest benefit is observed in high-spatial-resolution imaging protocols employing thin slice thicknesses and sharp reconstruction kernels, whereas the incremental benefit is limited for routine CT imaging. These findings support a task-based implementation of the UHR mode to optimize protocol selection according to clinical imaging requirements.
Abstract
BACKGROUND: Ultra-high-resolution (UHR) photon-counting CT (PCCT) with small-pixel detectors improves resolution and noise, but its dependence on diverse imaging conditions remains insufficiently characterized. PURPOSE: To systematically evaluate the small-pixel effect in UHR PCCT by assessing noise properties and high-contrast spatial resolution under varying acquisition and reconstruction conditions. METHODS: Images of an ACR CT accreditation phantom with a body ring simulating a medium-sized patient were acquired on a clinical PCCT system (NAEOTOM Alpha.Peak, Siemens Healthineers, Germany) using standard-resolution (STD) and UHR modes. Scans were performed at dose levels ranging from 3 to 24 mGy. Reconstructions included slice thicknesses from 0.2/0.4 to 3.0 mm, convolution kernels from Br36 to Br72, quantum iterative reconstruction both off and at strength levels 1-4, matrix sizes of 512 to 1024, and field-of-view (FOV) settings from 210 to 350 mm. Both low-energy threshold (T3D) and virtual monoenergetic images were evaluated. Image noise was quantified using noise magnitude, noise power spectrum (NPS), and average spatial frequency. High-contrast spatial resolution was evaluated using the modulation transfer function and visual assessment of resolution patterns. RESULTS: Compared with the STD mode, the UHR mode consistently reduced image noise, with the greatest reductions observed at thinner slice thicknesses and with sharper kernels. In contrast, dose level, QIR strength, matrix size, and spectral image type had minimal influence on the relative noise differences between the two modes. NPS analysis further demonstrated lower noise magnitude and, depending on the imaging conditions, a shift of the noise spectrum toward lower spatial frequencies, indicating relative suppression of high-frequency noise in the UHR mode. High-contrast spatial resolution was more strongly influenced by imaging sampling conditions, particularly matrix size and FOV, and remained overall comparable between the STD and UHR modes. CONCLUSIONS: The UHR mode of PCCT consistently achieves lower image noise than the STD mode without compromising high-contrast spatial resolution. Its greatest benefit is observed in high-spatial-resolution imaging protocols employing thin slice thicknesses and sharp reconstruction kernels, whereas the incremental benefit is limited for routine CT imaging. These findings support a task-based implementation of the UHR mode to optimize protocol selection according to clinical imaging requirements.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Small-pixel Acquisition Unmasks "Bucket Effects" of Iterative Reconstruction in Ultra-High-Resolution Photon-Counting CT.Academic radiology · 2026
- Acquisition Speed versus Spatial Resolution in Ultra-High-Resolution Photon-Counting CT: Phantom Study to Guide Protocol DevelopmentmedRxiv · 2026
- Quantitative and qualitative lung parameters in photon-counting CT: comparison of ultra-low-dose with standard low-dose protocols.European radiology · 2026
- OPTIMIZATION OF PHOTON-COUNTING CT FOR BONE IMAGING USING ULTRA-HIGH-RESOLUTION MODE.Osteoarthritis imaging · 2026
- Pediatric Chest Photon-counting CT: Dose and Noise Behavior of Clinical Sn100 kVp and 90 kVp Protocols across Image Quality Levels - a Phantom Study.RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin · 2026
- Ultra-high spatial resolution at photon-counting computed tomography: technical insights and sustainable applications in cardiothoracic imagingEuropean Radiology Experimental · 2026