Geometric decoupling of epidermal melanin and dermal hemoglobin signals using coaxial and spatially resolved NIRS.
Vedanshi Patel, Devang Vyas, John Hanks, Amir Tofighi Zavareh
Texas A&M University, Department of Biomedical Engineering, College Station, Texas, United States.
Key points
From the paper's conclusion- These findings support future experimental validation of integrated coaxial plus spatially resolved probes for melanin-robust absolute tissue oxygenation monitoring.
Abstract
SIGNIFICANCE: Epidermal melanin can confound NIRS measurements of tissue oxygen saturation and reduce accuracy across skin tones. AIM: We evaluated whether a coaxial detector can provide a superficial reference that improves absolute StO 2 estimation when combined with spatially resolved measurements. APPROACH: Monte Carlo simulations of multilayer skin were evaluated using mutual information and XGBoost regression, with entire melanin levels held out during validation. Three detector configurations were compared: coaxial only, surface only, and the combined geometry. A secondary experimental analysis examined absolute and baseline referenced oxygenation prediction using validation that excluded one recording at a time. RESULTS: The combined coaxial plus surface detector configuration reduced average StO 2 RMSE from 11.34 to 5.41 percentage points, corresponding to a 52.3% relative reduction based on the displayed fold results. Mutual information analysis showed that the coaxial detector carried substantial information about melanin while carrying negligible direct information about StO 2 . In the experimental analysis, baseline-referenced features improved pooled agreement, but the recording-based validation did not establish generalization to entirely unseen participants. CONCLUSIONS: These findings support future experimental validation of integrated coaxial plus spatially resolved probes for melanin-robust absolute tissue oxygenation monitoring.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Anatomical and spectral foundation of skin pigmentation bias in optical oximetrymedRxiv · 2026
- Investigate optical attenuation coefficient transformation associated with melasma in vivo by optical coherence tomography.Biomedical physics & engineering express · 2026
- Melanin- and linewidth-corrected pulse-oximetry.ArXiv · 2026
- Melanin Bias in Cerebral fNIRS: A Monte Carlo Study Toward Equitable Optical Brain Sensing.Biosensors · 2026
- Frequency-domain dynamic speckle dual-wavelength StO2 imaging: Feasibility of occlusion localizationApplied Physics Letters · 2026
- Physiological Parameter Estimation from Reduced-Wavelength Hyperspectral Data: A Model-Based StudyCurrent Directions in Biomedical Engineering · 2026