Evaluation of an ABO antibody immunoadsorption column for reduction of anti-A and anti-B isoagglutinin titers in group O platelet units.
Melanie J Robbins, Susan Proffitt, Gillian Eastwood, Sian Huish, Kurt Nilsson, Rebecca Cardigan
Component Development, NHS Blood and Transplant, Cambridge, UK. · GlycoRex Transplantation AB, Lund, Sweden.
Key points
From the paper's conclusion- Anti-A and anti-B titers were decreased from titers of 128 to ≤4 (IgM) and from 512 to ≤8 (IgG) following a single pass through the immunoadsorption column. Reduced efficacy was observed following a second pass through the same column (≤32 (IgM) and ≤128 (IgG)). Filtration was not associated with substantial platelet loss, although changes in some metabolic and activation markers were observed during storage.
Abstract
BACKGROUND: Clinical guidelines recommend transfusing ABO-identical platelets where possible. However, the short shelf life of platelets (5-7 days) and stock availability mean ABO-nonidentical platelet transfusions are sometimes unavoidable. Minor ABO-incompatible platelet transfusions carry a risk of hemolysis due to donor anti-A/B antibodies, and high-titer (HT)-negative units are preferred to mitigate this risk. To further mitigate this risk, we investigated whether an ABO antibody immunoadsorption column could reduce anti-A/B titers in group O platelet units while preserving platelet quality. STUDY DESIGN AND METHODS: Six group O double dose apheresis platelet units were pooled with HT-positive group O plasma and split into three arms: untreated control, single-pass filtration (Glycosorb ABO A/B-column), and double-pass using the same column. Anti A and anti B IgM and IgG titers were measured pre- and post-filtration using a gel card method. Platelet count, platelet volume, coagulation factors, and markers of platelet quality and activation were assessed after filtration and during storage. RESULTS: Anti-A and anti-B titers were decreased from titers of 128 to ≤4 (IgM) and from 512 to ≤8 (IgG) following a single pass through the immunoadsorption column. Reduced efficacy was observed following a second pass through the same column (≤32 (IgM) and ≤128 (IgG)). Filtration was not associated with substantial platelet loss, although changes in some metabolic and activation markers were observed during storage. DISCUSSION: The ABO antibody immunoadsorption column effectively reduced anti-A/B titers in group O platelet units and may offer a strategy to increase the availability of low-titer platelets for transfusion.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Novel In-line Filter Reduces ABO Circulating Antibodies in Whole Blood.Military medicine · 2026
- Covalent immobilization of blood typing antibodies on multilayer porous membranes for stable and rapid ABO blood typing immunodetection.Mikrochimica acta · 2026
- Phenotype-dependent effects of 2-mercaptoethanol pretreatment on ABO antibody reactivity by automated column agglutination.Frontiers in pharmacology · 2026
- Determination of the optimal pool size of ABO-matched buffy coat (BC) units to achieve a typical adult dose (TAD) of platelets comparable to single donor platelet concentrate (SDPC): A single-center study.Asian journal of transfusion science · 2026
- 16 Post-Storage Removal of the Supernatant in Platelet Units May Improve the Safety of Platelet Transfusions in Children: A Single Center ExperienceAmerican Journal of Clinical Pathology · 2026
- Maternal immunoglobulin G anti-A and anti-B antibody titres for predicting neonatal immune sensitisation in ABO-incompatible pregnancies: The Maternal Anti-A and Anti-B Antibody Titres in ABO-Incompatible Pregnancies study.Vox sanguinis · 2026