Novel In-line Filter Reduces ABO Circulating Antibodies in Whole Blood.
Gregory C Wetmore, Ryan C Chae, Lindsey J Wattley, Rebecca Schuster, Bernadin Joseph, Michael D Goodman, Jose A Cancelas, Maryann Gruda,
Department of Surgery, University of Cincinnati College of Medicine, Cincinnati, OH 45267-0558, United States. · CytoSorbents Corporation, Princeton, NJ 08540, United States.
Key points
From the paper's conclusion- The novel in-line filter effectively reduces circulating anti-A and -B antibody titers in whole blood, though the extent of reduction varies by infusion method. Filtration induces mild changes in hematologic and coagulation parameters, but overall, filtered whole blood maintains the critical attributes for hemostatic resuscitation. Further studies are warranted to optimize filtration rates for maximal antibody clearance and to assess the impact of filtration rates for maximal antibody clearance. This approach may expand the LTOWB donor pool by enabling safe use of high-titer anti-A/anti-B donors.
Abstract
INTRODUCTION: Low titer O whole blood (LTOWB) is the preferred blood product for resuscitation after hemorrhage in civilian and military trauma. The low titer requirement minimizes the risk of hemolytic transfusion reactions caused by donor ABO antibodies. However, up to 25% of otherwise eligible donors are excluded because of high antibody titers. A novel in-line filter has been developed to bind and remove donor antibodies from whole blood during infusion. We evaluated the efficacy of this filter across multiple infusion methods and assessed its effects on whole blood composition and coagulation potential. MATERIALS AND METHODS: This study was reviewed and approved as non-human research by the Institutional Review Board. Low-titer O whole blood (LTOWB) units were obtained from our regional blood bank. In the first experimental series, LTOWB units were split into unmodified native whole blood (n = 5) and spiked whole blood (n = 5) to which murine anti-human anti-A and -B antibodies were added to achieve elevated titers before simulated infusion using 3 commercially available methods (gravity, rapid manual, and automated rapid normothermic) and samples were collected after infusion both with and without the in-line filter. In a second series, unmodified native whole blood units were evaluated by cell count, flow cytometry, viscoelastic thromboelastometry, and platelet (PLT) aggregometry with and without in-line filtration. Comparisons were made using Mann-Whitney and paired t-tests. RESULTS: Use of the in-line filter significantly reduced anti-A and anti-B titers across all infusion methods, with the greatest reduction seen after gravity infusion. Gravity filtration was also associated with the most pronounced changes in whole blood characteristics, including reductions in white blood cell count, PLT count, hematocrit, and hemoglobin. Viscoelastic testing revealed modest alterations in maximum clot firmness, alpha angle, and PLT contribution to clot. Multiplate aggregometry demonstrated reductions in area under the curve on both arachidonic acid and adenosine diphosphate agonist assays. CONCLUSIONS: The novel in-line filter effectively reduces circulating anti-A and -B antibody titers in whole blood, though the extent of reduction varies by infusion method. Filtration induces mild changes in hematologic and coagulation parameters, but overall, filtered whole blood maintains the critical attributes for hemostatic resuscitation. Further studies are warranted to optimize filtration rates for maximal antibody clearance and to assess the impact of filtration rates for maximal antibody clearance. This approach may expand the LTOWB donor pool by enabling safe use of high-titer anti-A/anti-B donors.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Massive transfusion practices and low-titer group O whole blood utilization before and after implementation in a low-volume level I trauma center: A six-year review.Transfusion · 2026
- Evaluation of an ABO antibody immunoadsorption column for reduction of anti-A and anti-B isoagglutinin titers in group O platelet units.Transfusion · 2026
- Reducing Cell-Free Hemoglobin Levels in Freeze-dried Red Blood Cells for Transfusions in Austere Environments.Military medicine · 2026
- RhD-Positive Low-Titer O Whole Blood in Persons of Childbearing Potential: Known Risk, Unproven Benefit, and the Case for Restraint.American journal of obstetrics & gynecology MFM · 2026
- In vitro compatibility of whole blood with commonly used medications in the pediatric trauma setting.Hematology, transfusion and cell therapy · 2026
- Prehospital whole blood for trauma: when two trials agree, the burden of proof shifts.British journal of anaesthesia · 2026