Late Obesity-Driven Subcutaneous ICD Geometric Failure: Salvage With Sub-Serratus Generator Repositioning.
Federico Turreni, Pietro Francia, Maria Luisa Loricchio, Renzo Iulianella, Antonino Granatelli, Marco Testa
Division of Cardiology, Santa Maria della Stella Hospital, Orvieto, Italy. · Division of Cardiology, Department of Clinical and Molecular Medicine, Sant'Andrea Hospital, Sapienza University of Rome, Rome, Italy.
Key points
From the paper's conclusion- We report late obesity-associated geometric deterioration after substantial weight gain.
Abstract
S-ICD efficacy depends on extracardiac shock vector geometry. We report late obesity-associated geometric deterioration after substantial weight gain. Rising shock impedance prompted reassessment at generator replacement, revealing pocket dehiscence, adipose interposition, and generator superficialization. Because fibrosis precluded parasternal lead mobilization by simple traction, sub-serratus generator repositioning improved the PRAETORIAN score and shock impedance. Despite a subsequent rise to 130 Ω, defibrillation succeeded with a 65 J shock. S-ICD geometry should be considered dynamic, particularly after major changes in body habitus.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Reverse Ratchet Syndrome With Intracardiac Looping.Journal of arrhythmia · 2026
- Long-term outcomes of subcutaneous ICD therapy in arrhythmogenic cardiomyopathy: a multicentre cohort studyEuropean Heart Journal · 2026
- Long-term outcomes of subcutaneous implantable cardioverter defibrillators: the French nationwide HONEST studyEuropean Heart Journal · 2025
- Long term shock efficacy of the subcutaneous defibrillator in the EFFORTLESS extension sub studyEuropean Heart Journal · 2026
- Early Right Ventricular Implantable Cardioverter-Defibrillator Lead Microdislodgement Associated With Active-Fixation Helix Retraction: A Case Report.Cureus · 2026
- Implantation Technique, Continuous Implant Geometry, and Longitudinal Low-Voltage Impedance After Subcutaneous Implantable Cardioverter-Defibrillator Implantation.Journal of cardiovascular electrophysiology · 2026