Journal of microscopy

Accessible range of motion characterisation of the OpenFlexure microscope.

Ben Chisholm, Joe Knapper, Julian Stirling, William Wadsworth, Richard Bowman

School of Physics and Astronomy, University of Glasgow, Glasgow, Scotland. · Humanitarian Technology Trust, Bath, UK.

Computational

Key points

From the paper's conclusion
  • We present a method of measuring the range of motion of the OpenFlexure Microscope using minimal equipment and a computationally light algorithm suitable for running on embedded software in the microscope.

Abstract

For any hardware to remain accessible, calibration techniques which rely on low-cost and obtainable materials such as easily manufactured test slides are necessary. We present a method of measuring the range of motion of the OpenFlexure Microscope using minimal equipment and a computationally light algorithm suitable for running on embedded software in the microscope. The algorithm uses correlation to measure the displacement between consecutive images. The designed range of motion of the OpenFlexure Microscope is 12.5 mm . The method described measured the range of motion as 11.4 ± 0.4 × 11.4 ± 0.5 mm . This allows end-users to identify damage to the microscope and scan samples with confidence.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Accessible range of motion characterisation of the OpenFlexure microscope — Science Skim