Performance Evaluation of an Artificial Intelligence-Assisted Digital Morphology System for Bone Marrow Aspirate Examination.
S Urrutia Iñigo, J R Furundarena Salsamendi, A Uranga Aguirregomezcorta, L Agirrezabala Zurutuza, A Arambarri Oyarzabal, N Uresandi Iruin, N Argoitia Ituarte, L Urritza Mantiñan,
Hematology Laboratory, Donostia University Hospital, Donostia-San Sebastian, Spain. · Unit of Clinical Epidemiology and Research, Donostia University Hospital, Donostia-San Sebastian, Spain.
Key points
From the paper's conclusion- These findings, obtained in a single‐center series, provide a basis for further evaluation of this and similar systems.
Abstract
INTRODUCTION: Morphological examination of bone marrow aspirates (BMA) remains essential for the diagnosis of hematologic diseases, but conventional microscopy is labor-intensive and subject to interobserver variability. METHODS: A total of 158 samples from patients were analyzed. Differential counts obtained by optical microscopy were compared with those of the CellaVision BMA Application used according to its intended workflow of automated preclassification followed by expert review. The two methods were compared by Pearson correlation, Passing-Bablok regression, and Bland-Altman analysis, both overall and within the blast and plasma cells intervals that bracket the diagnostic thresholds. RESULTS: Pearson correlation coefficients ranged from 0.85 to 0.99 for all populations except basophils (r = 0.68). Passing-Bablok regression showed no significant systematic or proportional bias for erythroblasts, granulocytic cells, eosinophils, and monocytes; a small systematic difference was observed for blasts, a small proportional difference for lymphocytes, and significant deviations from linearity for plasma cells and basophils. Bland-Altman mean differences ranged from -1.4% to 1.2% across all populations. At the clinically relevant decision thresholds for blasts and plasma cells, overall agreement between the two methods ranged from 96.8% to 99.4%, although sensitivity at the 20% blast threshold was 77.3%. CONCLUSION: Used according to its intended workflow, the CellaVision BMA Application produced differential counts comparable with conventional microscopy for most nucleated cells, but agreement was weaker for basophils and plasma cells. Expert review and validation remain necessary. These findings, obtained in a single-center series, provide a basis for further evaluation of this and similar systems.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Beyond Analytical Agreement: Integrating Analytical Performance and Algorithm-Assisted Morphological Interpretation in a Comparative Evaluation of the Mindray BC-7800/BC-7900 and Sysmex XN-9100 Hematology Platforms.Diagnostics (Basel, Switzerland) · 2026
- Digital and computational morphology in hematology: current platforms, clinical evidence, and future requirements.Clinical chemistry and laboratory medicine · 2026
- AI-assisted digital peripheral blood morphology multicenter randomized paired method clinical validation study.NPJ digital medicine · 2026
- A-233 Clinical Validity of the AI-Based UIMD PBIA(Peripheral Blood Image Analyzer) in Leukopenic Samples (WBC <1.0×10^9/L)Clinical Chemistry · 2026
- Toward integrated multimodal artificial intelligence in bone marrow examination: current state and open challenges.Blood research · 2026
- Digital Morphology Analyzer-Based White Blood Cell Differentials: Practical Considerations for External Quality Assessment.Annals of laboratory medicine · 2026