The Contrast Order: An Order-Based Image Quality Criterion for Nonlinear Beamformers.
Dongwoon Hyun, Samuel Beuret, Jeremy J Dahl
Key points
From the paper's conclusion- Together, the CO and ECR provide order- and sign-preserving, dynamic-range-invariant criteria for evaluating lesion contrast, offering a principled alternative to classical and newer image quality criteria when assessing modern beamformers.
Abstract
Many modern ultrasound beamformers report improved image quality when evaluated using classical criteria like the contrast ratio and contrast-to-noise ratio, which are based on summary statistics of regions of interest (ROIs). However, nonlinear beamformers and post-processing methods can substantially alter these statistics, raising concerns that the reported improvements may reflect changes in dynamic range or remapping rather than a reflection of true information gain, such as clutter suppression. Introduced recently, the generalized contrast-to-noise ratio (gCNR) is designed to address these concerns, but it requires implementation-dependent density estimation. We propose herein a new, simpler, image quality criterion, called the contrast order (CO), defined as the expected value of the sign of the difference in brightness between two ROIs. The CO is invariant under all strictly monotonic transformations of the image values, as it depends only on their relative ordering, and is interpretable as the probability that one ROI is brighter than the other minus the probability that it is darker. Unlike the gCNR, the CO has a simple unbiased estimator whose variance provably decreases with the number of samples in each ROI. We further propose the effective contrast ratio (ECR), which calibrates the contrast order to the familiar contrast ratio such that the two coincide under ideal Rayleigh-speckle statistics. Together, the CO and ECR provide order- and sign-preserving, dynamic-range-invariant criteria for evaluating lesion contrast, offering a principled alternative to classical and newer image quality criteria when assessing modern beamformers.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Large-Scale Validation of Aberration Correction Algorithms in Clinical Fetal Ultrasound Scans.IEEE transactions on ultrasonics · 2026
- Task-based detectability and angular-sampling bounds on spatial resolution in cone-beam micro-CT: a closed-form scan-design analysis.Medical physics · 2026
- Lens-Aware Differentiable Beamforming for In Vivo Distributed Aberration Correction with Curvilinear Transducers.ArXiv · 2026
- CROSS-Calib: Selective Channel Calibration for Ultrasound ReconstructionarXiv · 2026
- Null Subtraction Imaging for functional ultrasound brain activation mapping.Ultrasonics · 2026
- A Method for Eliminating Mixed Errors in Confocal Images by Integrating System Error Matrices with Low-Rank ApproximationPhotonics · 2026