Adaptation of a Brain-Dedicated Multipinhole SPECT System for Imaging Higher-Energy Photons of Theranostic Agents-A Simulation Study.
Sophia Pells, Kesava S Kalluri, Micaehla May, Benjamin Auer, Lars R Furenlid, Matthew A Kupinski, Phillip H Kuo, Robert Licho,
Department of Radiology, University of Massachusetts Chan Medical School, Worcester, MA 01655 USA. · The University of Arizona James C Wyant College of Optical Sciences, Tucson, AZ 85721 USA.
Abstract
Multi-pinhole SPECT of the brain is capable of producing high-resolution images. However, pinhole imaging of high-energy photons remains challenging due to the trade-off of greatly sacrificing sensitivity to limit penetration through the edges of the pinholes and system shielding. The impact of photon penetration on spatial resolution, image contrast, and lesion signal-to-noise was assessed through simulation for the brain-dedicated multi-pinhole system AdaptiSPECT-C. In addition, minor changes to the system design to reduce penetration were considered; these were updating the materials of the pinhole aperture edges and shutters to a higher-density tungsten alloy and including platinum in the shielding around the pinhole apertures. A significant degradation in spatial resolution, image contrast, and signal-to-noise above approximately 200 keV was found with the current system design. It was shown that adapting the pinhole and shutter material to a higher-density tungsten alloy would greatly reduce this degradation and permit multi-pinhole imaging with 177Lu and 111In in the human brain. Incorporating a small amount of platinum around each aperture further improved the system point-spread function. However, results indicate high-resolution imaging of 131I would require a larger modification to the system design.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantitative Comparison of SPECT and PET Performance for Clinical Theranostic ApplicationsJournal of Nuclear Medicine · 2025
- Design and Development of NeuroSphere PET: A Next-Generation 7-T MR-Compatible Dedicated Brain PET Scanner.Journal of nuclear medicine : official publication, Society of Nuclear Medicine · 2026
- Impact of depth-of-interaction and spatially-variant point-spread-function model on the performance of ultra-high resolution brain PET.Physics in medicine and biology · 2026
- Characterization and Initial Optimization of Reconstruction Performance for Detector Crosstalk of Ultrahigh Performance Brain PET Imager.IEEE transactions on radiation and plasma medical sciences · 2026
- Beyond NEMA: robustness, DOI encoding, and multi-mouse imaging performance of the Bruker PET/CT Si78.Physics in medicine and biology · 2026
- A novel preclinical system for cascade gamma photon coincidence imaging with high resolution and sensitivity.Physics in medicine and biology · 2026