Department of Mechanical Engineering, Utah Robotics Center, The University of Utah, Salt Lake City, UT 84112 USA. · Department of Mechanical Engineering, Utah Robotics Center, and the Department of Biomedical Engineering, The University of Utah, Salt Lake City, UT 84112 USA, and also with the Rocky Mountain Center for Occupational and Environmental Health, Salt Lake City, UT 84111 USA.
Key points
From the paper's conclusion- Moreover, automatic transitions between gait cycles allow for seamless transitions between passive walking and active stair ambulation.
Abstract
The most commonly prescribed ankle/foot prostheses are energy storage and return (ESR) devices, which have good performance in level ground walking, but have limited functionality in tasks requiring ankle mobility, such as stair ascent or descent. Powered prostheses aim to address this limitation by actively controlling the joint with actuators, but they tend to be heavier and taller and become unusable when the battery is depleted, hindering their clinical adoption. In this paper, we present a semi-active ankle prosthesis that integrates the energy storage and return characteristics of a passive prosthesis with the active joint articulation of a powered prosthesis. The proposed prosthesis combines a novel brake mechanism with a series elastic actuator to enable Passive and Active operation modes within a compact build height (141 mm) and lightweight (1.4 kg without batteries) form factor. Experimental validation with one lower-limb amputee demonstrates that the proposed prosthesis reproduces the gait dynamics of a passive prosthesis, independent of battery charge, in Passive mode, and matches the range of motion and actuation capabilities of a powered device during stair ascent and sit-to-stand transitions in Active mode. Automatic transitions between modes within a gait cycle allow for incline adaptation. Moreover, automatic transitions between gait cycles allow for seamless transitions between passive walking and active stair ambulation. This combination of functionalities has never been demonstrated in an ankle prosthesis and has the potential to improve mobility among the amputee population.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Design and Initial Evaluation of a Low-Cost Microprocessor-Controlled Ankle ProsthesisSensors · 2026
- The variable stiffness orthosis: customizable mechanics for assistance and rehabilitationJournal of NeuroEngineering and Rehabilitation · 2026
- Effects of a Torsion-Adaptive Energy-Storing-and-Return Prosthetic Foot on Balance Control in Individuals with Transtibial Amputation: A Pilot Pre-Post Motion Analysis Study.Bioengineering (Basel, Switzerland) · 2026
- Design and Bench Evaluation of a Sensor-Integrated Prosthetic-Leg Prototype.Sensors (Basel, Switzerland) · 2026
- Crossing Obstacles during Walking through Self-adaptive Volitional Control of a Powered Knee Prosthesis.IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society · 2026
- On Ankle-Foot Prostheses: An Introduction to Biomechanical Goals, Modern Design Principles, Device Evaluation, and Current Clinical Practice.IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society · 2026