IEEE transactions on biomedical circuits and systems

A Wireless Neural-Interface SoC with Multimodal Chemical and Electrical Recording Capabilities.

Minghao Chen, Gangzheng Zhu, Hong Liao, Yuqian Li, Jing Huang, Yuanyu Tan, Bingfang Wang, Ying Fang,

Key points

From the paper's conclusion
  • Measurement results show DRs of 105.3 dB for FSCV and 138.8 dB for OECT.

Abstract

This article presents a wirelessly powered implantable System-on-Chip (SoC) for multimodal chemical and electrical neural signal monitoring. This is accomplished by: 1) proposing a low-power proportional-integral (PI) feedback loop that controls a background subtraction digital-to-analog converter (DAC) to track rapidly varying backgrounds and accelerate dynamic range (DR) extension; 2) employing an interpolating arbitrary waveform generator and reconfigurable potentiostat to provide low-ripple excitation across multiple modalities without off-chip components; 3) utilizing a rectifier-based dynamic voltage supply (DVS) that tracks electrode voltage requirements, significantly reducing high-voltage power loss. The platform supports fast-scan cyclic voltammetry (FSCV) with three-electrode sensors, local field potential (LFP) acquisition with organic electrochemical transistors (OECTs), and photoelectrochemical (PEC) sensors. Fabricated in a 65-nm complementary metal-oxide-semiconductor (CMOS) technology, the SoC occupies a core area of 0.33 mm2 and a peak power consumption of 86 μW. Measurement results show DRs of 105.3 dB for FSCV and 138.8 dB for OECT. The detected dopamine increase in striatum verified the functionality of the neural interface SoC in vivo.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A Wireless Neural-Interface SoC with Multimodal Chemical and Electrical Recording Capabilities — Science Skim