One-year evaluation of a 128-channel flexible IrOx neural interface and development of a miniaturized 256-channel integrated prototype.
Dongyang Wen, Longchun Wang, Guijun Shu, Yang Xiao, Linkun Gu, Xiao Wang, Ming Yin, Jingquan Liu
National Key Laboratory of Advanced Micro and Nano Manufacture Technology, Shanghai Jiao Tong University, Shanghai, 200240, China; School of Integrated Circuits (School of Information Science and Electronic Engineering), Shanghai Jiao Tong University, Shanghai, 200240, China. · National Key Laboratory of Advanced Micro and Nano Manufacture Technology, Shanghai Jiao Tong University, Shanghai, 200240, China.
Key points
From the paper's conclusion- Here, we combine a year-long evaluation of a 128-channel flexible IrOx electrode array (128-Flex) with the development of a miniaturized 256-channel integrated prototype (256-prot).
Abstract
High-density neural interfaces require both durable electrode-tissue interfaces and compact electronic integration. Here, we combine a year-long evaluation of a 128-channel flexible IrOx electrode array (128-Flex) with the development of a miniaturized 256-channel integrated prototype (256-prot). IrOx modification reduced the mean 1 kHz impedance from 1.096 MΩ to 18.15 kΩ and increased charge storage capacity from 2.02 to 15.37 mC/cm2. Periodic in vivo recordings extended to 12 and 7 months in two rabbits, respectively, with final-session mean SNR values of 6.18 and 5.88 among spike-detecting channels. Post-explantation characterization of the 12-month device showed retained electrode morphology and electrochemical functionality, with a mean impedance of 48.60 kΩ and charge storage capacity of 8.67 mC/cm2. Building on this electrode platform, two 128-channel ASIC modules were integrated through a resin-based feedthrough substrate and two BGA interfaces into a 13 × 13 × 3 mm bare module weighing 1.03 g. The prototype recorded LFPs and spike activity at 1.5 months in an independent rabbit, with 80 and 73 effective recording channels in separately acquired module datasets. Enclosure optimization remains a priority for extending integrated-system evaluation. These results support the long-term recording utility of the flexible IrOx interface and the feasibility of the compact integration strategy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Wireless Neural-Interface SoC with Multimodal Chemical and Electrical Recording Capabilities.IEEE transactions on biomedical circuits and systems · 2026
- Umbrella-type stretchable electrode arrays for long-term neural recording.Nature communications · 2026
- An Active, Multimodal Neural Interface for Real‐Time Monitoring of Cortical Electrical, Thermal, and Optical DynamicsAdvanced Science · 2025
- A Reconfigurable CMOS Microelectrode Array Featuring 24,576 Electrodes for Chemical and Electrical Neural Recording.IEEE transactions on bio-medical engineering · 2026
- Clinical translation of ultrasoft Fleuron probes for stable, high-density, and bidirectional brain interfacesmedRxiv · 2026
- Development and Innovative Applications of Tools for Monitoring Neuronal Electrical ActivityTheoretical and Natural Science · 2026