IEEE journal of translational engineering in health and medicine

Development and Bench-to-Human Feasibility Evaluation of a Wireless Pinch-and-Grip Dynamometer.

Ali Pourhossein, Mohamed Mehdi Ouerghi, Elisabeth Kuborn, Doguhan Tasdemir, Viktor Warwas, Charmayne Mary Lee Hughes

Faculty V-Mechanical Engineering and Transport Systems, Institute of Psychology and ErgonomicsTechnische Universität Berlin Berlin 10623 Germany.

Key points

From the paper's conclusion
  • The MeRKeL system enables accurate and repeatable, reference-calibrated wireless measurement of pinch and grip strength, establishing a robust technical foundation for subsequent clinical evaluation and future home-based neuromuscular monitoring studies.

Abstract

OBJECTIVE: This study presents MeRKeL, a portable wireless system designed for quantitative, reference-calibrated assessment of pinch and grip strength, and evaluates its technical validity and human feasibility under standardized laboratory conditions. METHODS: The system consists of two compact and functionally aligned modules (pinch and grip dynamometer) sharing a unified electronic architecture and data acquisition pipeline. Both employ task-specific mechanical designs tailored to the distinct loading characteristics of pinch and grip actions. Reference-based calibration and bench validation were performed across the full operating ranges and across multiple units. Accuracy, agreement, and within-session repeatability were evaluated using root mean squared error (RMSE), Bland-Altman analysis, and intraclass correlation coefficient (ICC). Feasibility testing in 196 healthy adults examined associations between maximal strength and demographic, behavioral, and health-related variables. RESULTS: Bench validation showed high agreement with the calibrated reference instrument (pinch: RMSE = 0.95 N, bias = -0.56 N; grip: RMSE = 1.86 N, bias = -1.10 N), and within-session repeatability was excellent (MDC95 pinch = 0.86 N; grip = 1.43 N). Experimental testing reproduced expected associations with key demographic and behavioral factors. CONCLUSION: The MeRKeL system enables accurate and repeatable, reference-calibrated wireless measurement of pinch and grip strength, establishing a robust technical foundation for subsequent clinical evaluation and future home-based neuromuscular monitoring studies.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers