Size and Attenuation Measurements for Urinary Stone on Virtual Monoenergetic Images Derived From Dual Layer Spectral CT: A Phantom Study.
Yannan Cheng, Yanfei Cheng, Yanan Li, Lu Zhang, Ganglian Fan, Xiaoyue Zhang, Huan Wang, Le Cao,
Departments of Radiology. · Neurology.
Key points
From the paper's conclusion- For VM images used to measure urinary stone size and attenuation, the optimal energy level required may differ based on the window settings. In addition, reconstructed thinner slice thickness can improve measurement accuracy.
Abstract
OBJECTIVE: This study aimed to investigate the optimal energy level for virtual monoenergetic (VM) images with specific window settings for stone size and attenuation measurements, and to determine the effects of radiation dose and reconstructed slice thickness on these measurements. METHODS: Forty-eight urinary calcium oxalate stones (<1 cm in diameter) were placed within a 32-cm phantom, and CT scans were performed on a dual-layer detector scanner at 3 radiation doses (volumetric computed tomography dose index=5, 10, and 15 mGy). Images were reconstructed as conventional 120-kVp and VM images ranging from 40 to 100 keV in 10-keV increments with a slice thickness of 1 mm. In addition, 70-keV images were reconstructed at slice thicknesses of 2 and 5 mm. Length, width, and attenuation were measured at bone and soft tissue window settings by 2 radiologists. The reference stone size was measured with a digital caliper, and the reference attenuation was measured on 120 kVp images. Statistical analyses were performed using the paired t test, the Wilcoxon test, and repeated-measures one-way analysis of variance. RESULTS: The stone size and attenuation decreased with increasing energy level of the VM images, irrespective of window settings. The stone size on the VM 70- to 90-keV images under the soft tissue window and on the VM 40-keV images under the bone window was closest to the true size. Attenuation values measured on the VM 70-keV images were similar to those obtained on the conventional 120-kVp images across both window settings. In addition, measurements of stone size (length and width) and attenuation on VM images with thinner slices were closer to the true value. CONCLUSIONS: For VM images used to measure urinary stone size and attenuation, the optimal energy level required may differ based on the window settings. In addition, reconstructed thinner slice thickness can improve measurement accuracy.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Image quality assessment of 40-keV virtual monoenergetic reconstruction in low-contrast lower-extremity spectral CT angiography.Radiography (London, England : 1995) · 2026
- Impact of acquisition parameters in photon-counting computed tomography on the stability of virtual monoenergetic attenuation values and effective atomic number: a phantom study.European journal of radiology · 2026
- Photon-Counting CT With Low-keV Virtual Monoenergetic Images for Detecting Muscle Invasion in Bladder Cancer: A Prospective Study.AJR. American journal of roentgenology · 2026
- Personalized low-iodine dual-layer spectral CT protocol for post-endovascular aortic repair imaging: a BMI/eGFR-adapted strategy preserving renal function, quality and diagnostic confidence.Clinical radiology · 2026
- Influence of X-ray Tube Voltage on Computed Tomography Values and Contrast-Noise Ratios in Virtual Monoenergetic Images in routine clinical Photon Counting Detector Computed Tomography: Does kV Still Matter?Investigative radiology · 2026
- Optimization of Posterior Fossa Image Quality Using Virtual Monoenergetic Reconstructions from Dual-Layer Spectral CT.Tomography (Ann Arbor, Mich.) · 2026