International journal of surgery case reports

Difficult INFIX screw removal after screwdriver fracture: salvage using an extracorporeal rod-screw complex.

Weidong Zhao, Rui Huang, Yun Zhou, Zheng Wang

Department of Orthopedic, General Hospital of Ningxia Medical University, Yinchuan, Ningxia Hui Autonomous Region, China.

Case report

Key points

From the paper's conclusion
  • When conventional methods fail, extracorporeal rod-screw complex assembly offers a bone-sparing salvage option for difficult INFIX screw removal. Preventive strategies - individualized screw sizing, preoperative instrument inspection, and timely implant removal - should be considered. Further experience with this technique is needed.

Abstract

INTRODUCTION: The INFIX (Anterior Subcutaneous Internal Pelvic Fixator) technique is widely used for minimally invasive stabilization of unstable pelvic ring injuries. Routine implant removal at 3-9 months postoperatively is usually straightforward, but difficult screw extraction can occur. Existing solutions are largely drawn from spinal pedicle screw retrieval techniques. CASE REPORT: A 26-year-old woman underwent elective removal of INFIX implants 9 months after open reduction and internal fixation for a Tile C1.1 pelvic fracture. Intraoperatively, the left pedicle screw was found to be tightly osseointegrated. When rotational force was increased, the screwdriver fractured, leaving the broken tip lodged within the screw recess. A new screwdriver could not engage the damaged head. After several retrieval attempts failed, we assembled an extracorporeal rod-screw complex by bending the connecting rod and locking it to the screw with the set screw, converting the polyaxial screw into a monoaxial construct. Using the rod as a lever, the screw was extracted without damaging the surrounding bone. At 6-month follow-up, the patient was pain-free, neurologically intact, and ambulating independently. CLINICAL DISCUSSION: The rod-screw complex principle was originally described for spinal pedicle screw removal. We applied it to the INFIX setting with three notable differences: extracorporeal assembly, rod contouring to avoid the anterior superior iliac spine, and application in the presence of a broken screwdriver tip retained within the screw recess - which, to our knowledge, has not previously been reported. Technical details, complications, and limitations are discussed. CONCLUSIONS: When conventional methods fail, extracorporeal rod-screw complex assembly offers a bone-sparing salvage option for difficult INFIX screw removal. Preventive strategies - individualized screw sizing, preoperative instrument inspection, and timely implant removal - should be considered. Further experience with this technique is needed.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Difficult INFIX screw removal after screwdriver fracture: salvage using an extracorporeal rod-screw complex — Science Skim