Comparing Oscillometric and Manual Mean Arterial Pressure Measurements Among Children in the Emergency Department.

Abidatou Diasso-Adamou, Cristina Payan, Danielle Cory, Sriram Ramgopal

Division of Emergency Medicine, Ann and Robert H. Lurie Children's Hospital.

Observational · n=347

Key points

From the paper's conclusion
  • Calculated oscillometric MAP had good agreement with device-reported oscillometric MAP, while manual MAP had poorer agreement with either oscillometric measure. Further study is needed to determine the clinical implications of these differences across age, weight, and patient-state subgroups.

Abstract

OBJECTIVE: To evaluate agreement among three approaches to mean arterial pressure (MAP) measurement in children. METHODS: We performed a single-center prospective study of a convenience sample of children (<18 y) presenting to a pediatric emergency department. We compared MAP agreement using the intraclass correlation coefficient (ICC) measured in three ways: calculated via auscultation, calculated from oscillometric systolic and diastolic measurements, and measurement via oscillometry. RESULTS: We included 347 patients (median age 8 y, IQR: 4 to 13). The highest ICC was observed between calculated and device-reported oscillometric MAP measurements (0.83, 95% CI: 0.79-0.86), indicating good agreement. Agreement was poor between manual and device-reported oscillometric MAP (ICC: 0.33, 95% CI: 0.22-0.43) and between manual and calculated oscillometric MAP (ICC: 0.38, 95% CI: 0.20-0.52). The median difference between calculated oscillometric and device-reported oscillometric MAP was 1.0 mm Hg (IQR: 0.0 to 2.3), compared with -3.3 mm Hg (IQR: -8.3 to 2.0) between manual and device-reported oscillometric MAP and -4.0 mm Hg (IQR: -9.3 to 0.3) between manual and calculated oscillometric MAP. Bland-Altman plots suggested that manual MAP measurements were slightly lower than both device-reported oscillometric and calculated oscillometric MAP measurements. Agreement between calculated oscillometric and device-reported oscillometric MAP was highest among children <2 years and lowest among children 12 to <18 years. Agreement involving manual MAP remained poor across most age groups. Similar patterns were observed when stratified by weight and patient state. CONCLUSIONS: Calculated oscillometric MAP had good agreement with device-reported oscillometric MAP, while manual MAP had poorer agreement with either oscillometric measure. Further study is needed to determine the clinical implications of these differences across age, weight, and patient-state subgroups.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Comparing Oscillometric and Manual Mean Arterial Pressure Measurements Among Children in the Emergency Department — Science Skim