Annals of medicine and surgery (2012)

Orbital volume measurement using direct measurement on 3D-printed models: a comparative study.

Mathilde De Massary, Pierre Labalette, Matthias Schlund, Benjamin Bouet, Emmanuel Chazard, Romain Nicot

Department of Ophthalmology, University Lille, CHU Lille, Service de d'Ophtalmologie, Lille, France. · Oral and Maxillofacial Department, University Bordeaux, CHU Bordeaux, INSERM, Inserm U1026 - Bioengineering of Tissues, Bordeaux, France.

Key points

From the paper's conclusion
  • The 3D-printing-assisted OV measurement method is reliable, reproducible, and cost-effective for orbital volumetric assessment in both clinical and preclinical experimental settings. This technique may serve as a valuable tool for validating computer-assisted OV measurement algorithms and for the in vitro evaluation of orbital implants and medical devices.

Abstract

BACKGROUND: Accurate orbital volume (OV) measurement is essential for the management of orbital fractures. However, no consensus exists regarding the optimal method for OV assessment. Manual computed tomography (CT) planimetry remains the most widely used technique in clinical practice. METHODS: This study aimed to develop a direct method for measuring OV using 3D-printed orbital models and to compare it with manual CT planimetry in healthy orbits. Eighteen healthy orbits were analyzed. OV measurements were obtained using two direct-filling techniques on 3D-printed models: sand filling (3D-Orb-1) and water filling (3D-Orb-2). The results were compared with manual CT planimetry (Plani-Orb). Reproducibility was assessed using intraclass correlation coefficients, and agreement between methods was evaluated using linear regression and Bland-Altman analysis. RESULTS: Mean OV values were 23.34 ± 2.15 cm3 for 3D-Orb-1, 23.31 ± 2.17 cm3 for 3D-Orb-2, and 24.10 ± 2.37 cm3 for Plani-Orb. Reproducibility was excellent for all methods, with ICC values of 0.963, 0.990, and 0.987 for 3D-Orb-1, 3D-Orb-2, and Plani-Orb, respectively. Strong correlations were observed between CT planimetry and both sand-filling (R 2 = 0.823; P < 0.0001) and water-filling techniques (R 2 = 0.798; P < 0.0001). Bland-Altman analysis demonstrated good agreement between methods. CONCLUSION: The 3D-printing-assisted OV measurement method is reliable, reproducible, and cost-effective for orbital volumetric assessment in both clinical and preclinical experimental settings. This technique may serve as a valuable tool for validating computer-assisted OV measurement algorithms and for the in vitro evaluation of orbital implants and medical devices.

Read the paper